Product specifications
96
Operating Manual for UTD2000/3000
DC measurement precision (by adopting
average sampling mode)
Measurement precision of voltage
difference (△V) (by adopting sampling
mode of average value)
±【(3% (reading+vertical displacement reading)+(1% vertical
displacement reading)】;
To add 2mV when setting from 2mV/div to 200mV/div: To add set
value by 50mV when setting from 200mV/div to 5V/div;
UTD2000L:
If vertical displacement is 0 and N≥16:
±(5%×reading+0.1 grid+1mV) by selecting 1mV/div or 2mV/div;
±(4%×reading+0.1 grid+1mV) by selecting 5mV/div;
±(3%×reading+0.1 grid+1mV) by selecting 10mV/div to 20V/div;
If vertical displacement is not 0 and N≥16:
±【(3%×(reading+vertical displacement reading)+(1%×vertical
displacement reading)】+0.2div;
To add 2mV when setting from 5mV/div to 200mV/div: To add set
value by 50mV when setting from 200mV/div to 20V/div;
××
Voltage difference between any 2 points on the waveform after
calculating average value for at least 16 captured waveform under the
same setup and environment conditions: ±(3%×reading+0.05div)
Note: Deflection coefficient scope of UTD2025C/3025C is 2mV/div ~ 10V/div (in input BNC).