Product specifications

92
Technical Indicators
Chapter 6: Appendix
Appendix A: Technical Indicators
Unless otherwise specified, all technical dimensions
are appropriate to probes by setting attenuation switch
to 10× and UTD2000/3000 series digital storage
oscillograph. Digital storage oscillograph must meet 2
conditions as follows to meet such dimension
standards:
Instrument must operate under regulated
operation temperature for more than 30 minutes
Note: Real-time sampling rate is 250MS/s without
continuously.
equivalent sampling for UTD2025C, UTD3025C and
Be sure to open system function menu and
UTD2025CL.
execute self-correctionprocedure if variation
scope of operating temperature reaches or
exceeds 5℃. All dimensions shall be guaranteed
except dimension with “typical” pattern.
Operating Manual for UTD2000/3000
Sampling
Sampling mode
Real-time
Equivalent
Sampling
rate
UTD2000C/3000C
UTD2000E/3000E
UTD2052CL
UTD2052CEL/
UTD2102CEL
500MS/s
500MS/s
1GS/s
1GS/s
25GS/s
25GS/s
50GS/s
50GS/s
Average
value
N can be selected from 2, 4, 8, 16, 32, 64,
128 and 256 after sampling all channels for
N times at the same time.