Product specifications
90
mode shall be used for video signals. Waveform can average times exceed 3 2.
not be displayed stably until setting correct trigger ② To reduce average times.
mode.
7. Staircase waveform display
③ Please strive to set coupling to high-frequency
① As normal phenomenon, horizontal time base
inhibition and low-frequency inhibition display to
can increase under super-low horizontal time
filter high/low-frequency noise triggered by
base gear so as to increase horizontal resolution
interference.
and improve display.
5. To press RUN/STOP key without any display:
② Possible display type is vector and sampling
① To check if trigger mode of trigger menu is in
points can be connected for staircase waveform
normal or single gear or not. It has to check if trigger
display. Display type can be set to point display.
level is out of waveform scope or not.
Trigger level shall be in the middle or trigger mode
shall be set to AUTO gear if yes.
② To press AUTO key for automatic setup as
above.
6. Display speed will reduce after opening average
time sampling mode.
①
In general, speed will reduce normally if
Operating Manual for UTD2000/3000