User Manual

86
UTD1000L
User Manual
Making Cursor Measurements
To make cursor measurements, do the following:
1. Press SHIFT and shift icon shows on upper
right part of display.
2. Press CURSOR to access cursor menu
3. Press F2 of UTD1000DL to choose the
channel for cursor measurement (No selection
for UTD1000CL, it defaults at CH1)
4 Press F1 to set up measuring type.
There are two options available for measuring
Type in cursor menu: Voltage and Time.
When measuring voltage, use SELECT and
arrow buttons to adjust the locations of two
cursors in order to measure parameter
do the same with measurement
Cursor 1 and
Cursor 2 shows simultaneously on the screen
us
e arrow buttons to remove the cursors and
press SELECT to choose and adjust the desired
one Displayed readout is the voltage or time
between two cursors
When the cursor function turns on, measuring
values can be automatically shown on upper left
corner of the display.
Figure 6-28 Cursor Menu for Time Measurement
;
.
,
.
:
,
.
.
ΔV
Δt
Voltage/Time Measurement Tips