User Manual
UTD1000L
User Manual
51
Data Hold Making Relative Measurement
You can freeze the displayed readings any time as
you want.
1 Press RUN /STOP to freeze measuring values
and HOLD icon blinks on the display
2 Repress RUN /STOP to restore the
measurement.
Figure 5-21 Freezing Measurement Values
Relative measurement displays a current
measuring result relative to the specific reference
value.
Refer to capacitance measurement in relative
mode. First of all you need to obtain a reference
value:
1. Press button to set measurement type to
Resistance
2 Press F1 button to select Capacitance option
3. Connect test leads to input terminals (Black to
COM and Red to V/
4. Wait until the reading becomes stable, then
press F2 button to access the relative mode,
icon shows on the upper display and the
reference value appears just below icon.
5. Connect test leads to the capacitor under test,
then take capacitance reading on the display
. ,
.
.
.
.
)
△
△
.
R
Ω