User Manual

UTD1000L
User Manual
17
then connect the probe connector to the
oscilloscope input channel and the probe tip to
compensation output terminal
2. Then execute AUTO
3. Observe the displayed waveform
If the displayed waveform is under-or over-
compensated as indicated above, use the adjusting
tool with non-metal handle to adjust the variable
capacitor on the probe until correctly compensated
waveform shows
To avoid any electric shock when
measuring the high voltage with use of the probe,
please ensure the probe wire is perfectly insulated,
and do not touch any metal part of the probe while
connecting to the high volt
age source.
Press V mV buttons to adjust the vertical
scale(VOLT/DIV) and display waveform with proper
size on the screen
When channel cursor turns solid press Up
and Down button to adjust ground reference
point that can allow the waveform to display at
proper location
;
.
;
,
;
Warning:
Vertical System
Setting Vertical&Horizontal Systems
and Trigger Level Manually
Over compensation
correct compensation
under compensation