User Manual
9
8
UTD1000L
User Manual
type has been set up can the waveform display
steady.
3. Try to change Coupling to HF Reject to filter out
the high-frequency inteference noise in the
signal.
1 Check if Trigger Mode in trigger menu is
Normal or Single and trigger level has gone
beyond waveform ranges. If they are, set trigger
level to the middle or set trigger Mode to
AUTO
2. Or you can press AUTO to automatically finish
the setup mentioned above
1. If AVG number go above 32times, it is normal for
the oscilloscopes to display at a relatively low
rate.
2. Reduce AVG number
1. It is normal situation, it may be caused by lower
horizontal ti
mebase, increasing the timebase
can improve the horizontal resolution and the
display
2. If display type is VECTOR ,links between sample
points could probably result in stair-stepping
shape. Change display type to Dots can solve
the problem.
There is no display after pressing
down RUN STOP button
The display slows down after having
applied AVG mode for a while:
The waveform display has a stair-
stepping shape:
/ :
.
.