Specifications

info@tursdaletechnicalservices.co.uk
20
取下電池蓋: Remove the battery compartment
取下產品後殼: Remove the case bottom
按需要更換保險絲: Install fuses with correct type and specification
Figure 5-1. Fuse Replacement
Warning
To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
Follow Figure 5-1 and proceed as follows to replace the Meter’s fuse:
z Turn the rotary switch to OFF and remove all connections from the terminals.
z Remove the screw from the battery compartment, and separate the battery compartment from the
case bottom.
z Remove the three screws from the case bottom, and separate the case bottom from the case top.
z Remove the fuse by gently prying one end loose, then take out the fuse from its bracket.
z Install ONLY replacement fuses with the identical type and specification as follows and make sure
the fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 250V, fast type fuse, ø5×20mm
Fuse 2: 10A, 250V, fast type fuse, ø5×20mm
z Rejoin the case bottom and case top, battery compartment and case bottom, and install the 5
screws.
Replacement of the fuses is seldom required. Burning of a fuse always results from improper operation.
C. Replacing the Battery
Figure 5-2. Battery Replacement