Model UT81B OPERATING MANUAL
Model UT81B: OPERATING MANUAL Chapter Title 1 Before You Start Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols Using The Testing Tool Reading the screen The Meter Structure Functional Buttons Making Measurements A. Scope Mode i. Setting up the Sleep Mode, Contrast, Beep ii. ACV, DCV, Hz, ACV and DCA Range iii. Trigger function iv. Waveform data save and recall B. Digital Multimeter Mode i. Measuring Voltages ii. Measuring Currents iii.
Model UT81B: OPERATING MANUAL Chapter 4 5 6 2 Title vii. Measuring Capacitance Using Software Maintenance The Test Tool A. General Service B. Replacing the Fuses C. Replacing the Battery Specifications Safety and Compliances Physical Specifications General Specifications (Digital Multimeter) General Specifications (Scope) Feature Summary Basic Specifications (Digital Multimeter) Basic Specifications (Scope) Detailed Accuracy Specifications A. DC Voltage B. AC Voltage C. DC Current D. AC Current E.
Model UT81B: OPERATING MANUAL Chapter Title H. Frequency and Duty Cycle % I.
Model UT81B: OPERATING MANUAL 4 Table Title 1-1 1-2 2-1 2-2 Unpacking Inspection International Electrical Symbols Reading the Screen Functional Buttons List of Tables Page 7 10 11 13
Model UT81B: OPERATING MANUAL Figure Title 2-1 2-2 3-1 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3-9 3-10 5-1 5-2 Meter Structure Functional Buttons Waveform Display Voltages Measurement µA Range Measurement mA Range Measurement 10A Range Measurement Resistance Measurement Diode Test Continuity Test Frequency / Duty Cycle Measurement Capacitance Measurement Fuse Replacement Battery Replacement List of Tables Page 12 13 17 21 23 24 25 27 29 31 33 35 39 40 5
Model UT81B: OPERATING MANUAL Chapter 1 Before You Start Overview This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly. bandwidth 8MHz, real-time sample rate 40MS/s with peak rate sampling process can catch up pulse industrial signal. It can measure AC/DC engine, transducer, circuit, control, UPS and industrial equipments. It is an ideal tools in professional repairing industries.
Model UT81B: OPERATING MANUAL Unpacking Inspection Open the package case and take out the Meter. Check the items shown on Table 1-1 carefully to see any missing or damaged part: Table 1-1. Unpacking Inspection Item Description Qty 1 English Operating Manual 1 piece 2 USB interface cable 1 piece 3 CD-ROM (Installation Guide & Computer Interface Software) 1 piece 4 Test Lead 1 pair 5 Alligator Clip 1 piece 6 Power Adaptor 1 piece 7 1.
Model UT81B: OPERATING MANUAL Safety Information This Meter complies with the standards IEC61010 safety measurement requirement: in pollution degree 2, overvoltage category (CAT. II 1000V, CAT.III 600V) and double insulation. International electrical symbols used on the Meter and in this Operating Manual are explained on page 10. Rules For Safe Operation Warning CAT.II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient overvoltages than CAT. III CAT.
Model UT81B: OPERATING MANUAL shock or damages to the Meter. l The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter. l When the Meter working at an effective voltage over 60V in DC or 42V rms in AC, special care should be taken for there is danger of electric shock. l Use the proper terminals, function, and range for your measurements.
Model UT81B: OPERATING MANUAL International Electrical Symbols Symbols used on the Meter and in this manual are explained in Table1-2. Table 1-2. International Electrical Symbols AC or DC DC Measurement AC Measurement Continuity Test Diode Grounding Double Insulated Warning.
Model UT81B: OPERATING MANUAL Chapter 2 Using the Testing Tool Reading the Screen The screen displays the menu that provides the following choices available: Table 2-1. Reading the Screen Display Contrast Auto Off BK Light BEEP ENTER MOVE MOVE RANG Description The degree of contrast Sleep mode time Display backlight Beeper on and off Confirm Increase Decrease Waveform moves up Waveform moves down Increase a range Table 2-1.
Model UT81B: OPERATING MANUAL The Meter Structure The Figure 2-1 shows the Meter structure. 1. 2. 3. 4. 5. 6. 7. 8. 9. USB Terminals LCD Display Functional Buttons Rotary Switch Power adaptor Input Terminals 10A Input Terminal mAµA Input Terminals COM Input Terminal Other Input Terminals Figure 2-1.
Model UT81B: OPERATING MANUAL Functional Buttons The buttons activate features that augment the function selected with the rotary switch. The buttons are shown in Table 2-2. Table 2-2. Functional Buttons Buttons F1, F2, F3 and F4 Range Time Trig Auto Figure 2-2. Functional Buttons Description Software functional buttons, details please refer to the below. Under scope mode, Press Range button to switch between DC and AC measurement Under scope mode, press Time button to set the X-axis of time base.
Model UT81B: OPERATING MANUAL Table 2-2. Functional Buttons Description To switch between waveform display (scope mode) and digital reading (multimeter mode). This button is only valid when under voltage, frequency, currents mode. Press Set button to set the auto power Set off, backlight, contrast and beep Save/Call Under scope mode, press Save/Cal to store and recall data. Press Hold button to enter or exit hold Hold mode.
Model UT81B: OPERATING MANUAL Chapter 3 Making Measurement Introduction Chapter 3 explains how to make measurements. You could turn the Meter off by turning to OFF position or set up the sleep mode from 1-30 minutes. Please must ensure the Meter is not under sleep mode if you turn the Meter on but without display. To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ” appears. A.
Model UT81B: OPERATING MANUAL F2: Set the Display Backlight BK Light 15 ENTER F1 F2 F3 F4 The brightness level from 0 to 31. Press F4 to confirm, save and return. Press functional buttons to exit, the setting is kept, but will not save. The setting will be lost after power off. F3: Set the LCD contrast Contrast 15 ENTER F1 F2 F3 F4 The contrast level from 0 to 31. Press F4 to confirm, save and return. Press functional buttons to exit, the setting is kept, but will not save.
Model UT81B: OPERATING MANUAL Turn the rotary switch to ACV, DCV, Hz, ACA or DCA range, the Meter displays digital reading (Multimeter mode). Press Mode to switch to waveform display (scope mode) as below Figure 4. When entering scope mode, time base is auto trace, the amplitude is manual set, you may need to re-set them. You could set the trigger level as well if it is needed. Details of measurement operation of ACV, DC, Hz, ACA or DCA can be seen from B.
Model UT81B: OPERATING MANUAL l Y-axis adjustment: Press Range button under scope mode, the corresponding functional button: Move Move Rang Rang F1 F2 F3 F4 F1: F2: F3: F4: move up the waveform move down the waveform go up range go down a range The auto set feature will be off when changing the measurement mode l Press Time button under scope mode, the corresponding functional button: Base Base Base < Base > F1 F2 F3 F4 18 F1: F2: F3: F4: increase the number of periods decrease the number of periods.
Model UT81B: OPERATING MANUAL Press Trig button under scope mode, the corresponding function buttons: Trig Trig Auto/Norm/Shot Slop Rrise/Fall F1 F2 F3 F4 F1: F2: F3: F4: move the trigger level up move the trigger level down select the trigger mode: auto, normal or single slope adjustment: rise or fall iv.
Model UT81B: OPERATING MANUAL be returned to working mode to carry out measurement. In order to have more accurate waveform, user can buy an optional BNC probe and scope probe to decrease signal attenuates. The scope probe directly connect to the BNC probe. When measuring voltage and frequency signal, connect the BNC black probe to the COM input terminal and the red probe to the voltage terminal. When measuring current signal, connect BNC black probe to the COM terminal and the red probe to mA terminal.
Model UT81B: OPERATING MANUAL When measuring voltage, the corresponding functional buttons AC / DC REL Rang Rang F1 F2 F3 F4 F1: F2: toggle between AC or DC relative mode (REL will be displayed at the right bottom of the LCD when it is on) select a range up select a range down F3: F4: Note: l l After changing the measurement mode, the autoranging will be off automatically and the AUTO will be disappeared at the bottom left of the LCD.
Model UT81B: OPERATING MANUAL Warning If the fuse burns out during measurement, the Meter may be damaged or the operator himself may be hurt. To avoid possible damage to the Meter or to the equipment under test, check the Meter’s fuses before measuring current. Use proper terminals, function, and range for the measurement. Never place the testing leads in parallel with any circuit or component when the leads are plugged into the current terminals.
Model UT81B: OPERATING MANUAL l To measure ACµA or DCµA currents, set up the Meter as Figure 3-3 and proceed as follows: 1. Insert the red test lead into the µmA terminal and black test lead into the COM terminal. 2. Set the rotary switch to µA . 3. Connect the test lead in series with the return circuit to be tested. 4. The measured value shows on the display. 5. Press MODE button to toggle between Multimeter mode and Scope mode. 6. Press F1 to toggle between AC and DC current measurement. Figure 3-3.
Model UT81B: OPERATING MANUAL l To measure ACmA or DcmA currents, set up the Meter as Figure 3-4 and proceed as follows: Figure 3-4. mA Range Measurement 24 1. Insert the red test lead into the µAmA terminal and black test lead into the COM terminal. 2. Set the rotary switch to mA . 3. Connect the test lead in series with the return circuit to be tested. 4. The measured value shows on the display. 5. Press MODE button to toggle between Multimeter mode and Scope mode. 6.
Model UT81B: OPERATING MANUAL l To measure AC 10A or DC 10A currents, set up the Meter as Figure 3-5 and proceed as follows: 1. Insert the red test lead into the 10A terminal and black test lead into the COM terminal. 2. Set the rotary switch to A . 3. Connect the test lead in series with the return circuit to be tested. 4. The measured value shows on the display. 5. Press MODE button to toggle between Multimeter mode and Scope mode. 6. Press F1 to toggle between AC and DC current measurement.
Model UT81B: OPERATING MANUAL When measuring current, the corresponding functional buttons: AC / DC F1 F1: F2: F3: F4: 26 REL F2 Rang F3 Rang F4 toggle between AC or DC relative mode (REL will be displayed at the right bottom of the LCD when it is on) select a range up select a range down Note l After changing the measurement mode, the autoranging will be off automatically and the AUTO will be disappeared at the bottom left of the LCD.
Model UT81B: OPERATING MANUAL Warning To avoid possible damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance. To measure resistance, set up the Meter as shown in Figure 3-6 and follow the following procedure: Figure 3-6.
Model UT81B: OPERATING MANUAL 1. Insert the red test lead into the Ω terminal and the black test lead into the COM terminal. 2. Set the rotary switch to Ω 3. Connect the test leads across with the object being measured. 4. The measured value shows on the display.
Model UT81B: OPERATING MANUAL To test the diode out of a circuit, set up the Meter as Figure 3-7 and proceed as follows: Warning To avoid harms to you, please do not attempt to input voltages higher than 60V DC or 42V rms AC. To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before testing diodes. Use the diode test to check diodes, transistors, and other semiconductor devices.
Model UT81B: OPERATING MANUAL 1. Insert the red test lead into the Ω terminal and the black test lead into the COM terminal. 2. Set the rotary switch to Ω . 3. For forward voltage drop readings on any semiconductor component, place the red test lead on the component’s anode and place the black test lead on the component’s cathode. The red test lead polarity is “+” while the black test lead polarity is “— “. The measured value shows on the display.
Model UT81B: OPERATING MANUAL To test for continuity, set up the Meter as Figure 3-8 and do the following: Warning To avoid harms to you, please do not attempt to input voltage higher than 60V DC or 42V rms AC. To avoid possible damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring continuity. Figure 3-8.
Model UT81B: OPERATING MANUAL 1. Insert the red test lead into the W terminal and the black test lead into the COM terminal. 2. Set the rotary switch to Ω . 3. Connect the test leads across with the object being tested. 4. The tested circuit overload resistance value shows on the display. 5. The beeper comes on continuously for open conditions, that is test resistance 10W. 6.
Model UT81B: OPERATING MANUAL Warning To avoid harms to you, please do not attempt to input voltage higher than 42V rms. To measure frequency and duty cycle, connect the Meter as Figure 3-9 and do the following: 1. Insert the red test lead into the Hz terminal and the black test lead into the COM terminal. 2. Set the rotary switch to Hz. 3. Connect the test leads across with the object being measured. 4. The measured value shows on the display. 5.
Model UT81B: OPERATING MANUAL Note l The requirement of Input amplitude “a” is as follows: When 1MHz: 300 mV a 30Vrms; 1MHz: 600 mV a 5Vrms l It is normal to have few seconds run time when switch from other functions to these functions. l When Hz or duty cycle measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove the test leads away from the input terminals.
Model UT81B: OPERATING MANUAL To measure capacitance, set up the Meter as shown in Figure 3-10 and proceed as follows: Warning To ensure accuracy, the Meter inside is discharged against the tested capacitor To avoid damage to the Meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Figure 3-10.
Model UT81B: OPERATING MANUAL 1. Insert the red test lead into theterminal and the black test lead into the COM terminal. 2. Set the rotary switch to measurement mode, the Meter may display a fixed reading which is a internal distributed capacitor value. For testing less than 40nF capacitor, the tested value must subtract the internal distributed capacitor value to maintain the accuracy. 3.
Model UT81B: OPERATING MANUAL Chapter 4 Using Software When using the Software, please refer to the Installation Guide of the included CD-ROM.
Model UT81B: OPERATING MANUAL Chapter 5 Maintaining The Test Tool This chapter provides basic maintenance information including battery and fuse replacement instruction. Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. A. General Service l Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
Model UT81B: OPERATING MANUAL B. Replacing the Fuses 1.Remove the 2.Remove the battery compartment case bottom Follow Figure 5-1 and proceed as follows to replace the Meter’s fuse: 3.Install fuses with correct type and specification Figure 5-1. Fuse Replacement Warning To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure. l Turn the rotary switch to OFF and remove all connections from the terminals.
Model UT81B: OPERATING MANUAL C. Replacing the Battery Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ” appears. Make sure the test leads are disconnected from the circuit being tested before opening the case bottom. Follow Figure 5-2 and proceed as follows to replace the battery: l Turn the rotary switch to OFF and remove all connections from the terminals.
Model UT81B: OPERATING MANUAL Chapter 6 Specifications Safety and Compliances Maximum Voltage between any Terminal and Grounding Certification Compliances Fused Protection for mA input terminal: Fused Protection for 10A input terminal: Refer to different range input protection voltage IEC 61010 CAT. II 1000V, CAT.III 600V overvoltage and double insulation standard 0.
Model UT81B: OPERATING MANUAL Physical Specifications Display (LCD) Operating Temperature Storage Temperature Relative Humidity Altitude Battery Type Electromagnetic Compatibility Dimensions (H x W x L) Weight 42 Digital: 3999 counts on display ; updates 2-3 times / second. o o o o 0 C~40 C (32 F~104 F) o o o o -10 C~50 C (14 F~122 F) o o 75% @ 0 C~40 C; o o 0% @ -10 C~50 C. Operating: 2000m; Storage: 10000m. 1.5V (R6) x 4 Batteries or Power adaptor.
Model UT81B: OPERATING MANUAL General Specifications ( Digital Multimeter) Range Polarity Overloading Battery Deficiency When it is under Multimeter mode, you could select either auto or manual ranging.
Model UT81B: OPERATING MANUAL Feature Summary Display Autorange Continuity Duty Cycle Battery Access Door 160 x 160 Monochrome When it is under multimeter mode, the Meter automatically selects best range Beeper sounds for resistance readings below threshold. Measure signal on or off time in %. Battery replaceable.
Model UT81B: OPERATING MANUAL Basic Specifications (Scope) Horizontal Sampling rate Sampling rate / Scale Updating rate Trigger types Timebase Range Timebase accuracy 40M per second 20 pixels >5 Free Run / Normal / Single Shot 100ns/div~5 sec /div (1-2-5) (0.
Model UT81B: OPERATING MANUAL A. DC Voltage Range 400mV 4V 40V 400V 1000V Resolution 100µV 1mV 10mV 100mV 1V Accuracy (0.8%+8) Overload Protection 1000V DC or AC Input Impedance Around 10MΩ (excluding waveform) (0.1%+8) B. AC Voltage Range 4V 40V 400V 750V Resolution 1mV 10mV 100mV 1V Accuracy (1%+15) Overload Protection 1000V DC or AC (1.2%+15) Remarks: l Frequency Response: 40Hz ~ 400Hz l Displays effective value of sine wave (mean value response).
Model UT81B: OPERATING MANUAL l The specification is same as Bi Under Multimeter mode. l On the top right corner of the display shows the accuracy of the frequency which is around 2%. When the LCD displays 1-10 period waveform, the reading is valid, otherwise the Meter displays “ —— “ C. DC Current Range Resolution 400µA 4000µA 40mA 400mA 4A 10A 0.1µA 1µA 10µA 100µA 1mA 10mA Accuracy (1%+8) Overload Protection 0.5A, 250V, fast type fuse, ø5×20mm (1.2%+8) (1.5%+8) 10A, 250V, fast type fuse, ø5×20mm.
Model UT81B: OPERATING MANUAL D. AC Current Range Resolution 400µA 4000µA 40mA 400mA 4A 10A 0.1µA 1 µA 10µA 100µA 1mA 10mA Accuracy (1.5%+8) 0.5A, 250V, fast type fuse, ø5×20mm (2%+8) (2.5%+5) Remarks: l Frequency Response: 40Hz~400Hz. l Displays effective value of sine wave (mean value response). 48 Overload Protection 10A, 250V, fast type fuse, ø5×20mm (Continuous measurement 10 seconds and interval more than 15 minutes.
Model UT81B: OPERATING MANUAL l The specification is same as Di Under Multimeter mode. On the top right corner of the display shows the accuracy of the frequency which is around displays 1-10 period waveform, the reading is valid, otherwise the Meter displays “------ “ 2%. When the LCD E. Resistance Range Resolution 400Ω 4kΩ 40kΩ 400kΩ 4MΩ 40MΩ 0.1Ω 1Ω 10Ω 100Ω 1kΩ 10kΩ Accuracy Overload Protection (1.2%+5) (1%+5) 250VDC or AC rms (1.2%+5) (1.5%+5) F.
Model UT81B: OPERATING MANUAL G. Continuity Test Test Range Resolution 0.1Ω Overload Protection 250V DC or AC Remarks l The buzzer sounds when the test resistance is 10Ω. l The buzzer does not sound when the test resistance is 100Ω. H. Frequency and Duty Cycle % Range Resolution Accuracy Overload Protection 10Hz~10MHz 0.1%~99.9% 0.001Hz 0.1% (0.
Model UT81B: OPERATING MANUAL l The specification is same as Hi. Under Multimeter mode, press RANGE button to toggle between frequency and duty cycle mode. l On the top right corner of the display shows the waveform’s RMS value, for reference only. When the LCD displays 1-10 period waveform cycle or the peak-peak difference value is greater than 1/3 div, the reading is valid, otherwise the Meter displays “ ————“ I.
Model UT81B: OPERATING MANUAL ** END ** This operating manual is subject to change without notice.
Model UT81B: OPERATING MANUAL Copyright 2006 Uni-Trend Group Limited. All rights reserved. Manufacturer: Uni-Trend Technology (Dongguan) Limited Dong Fang Da Dao Bei Shan Dong Fang Industrial Development District Hu Men Town, Dongguan City Guang Dong Province China Postal Code: 523 925 Headquarters: Uni-Trend Group Limited Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email:info@uni-trend.com http://www.uni-trend.
Model UT81B: OPERATING MANUAL 54