Model UT713 OPERATING MANUAL
Model UT713: OPERATING MANUAL TABLE OF CONTENTS TITLE Introduction Unpacking Inspection Safety Information Turning the Calibrator On Simulating a Thermocouple Measuring a Thermocouple Simulating TC or Source mV Measure TC or mV Cold Terminal Temperature Autocompensation Explanation of International Symbols Maintenance Specifications Temperature Measure and Thermocouple Simulate PAGE 3 3 4 6 6 7 8 10 12 14 15 19 19 1
Model UT713: OPERATING MANUAL TABLE OF CONTENTS TITLE Thermocouple Standards and Scales Millivolt Measure and Source General Specifications 2 PAGE 21 21 22
Model UT713: OPERATING MANUAL Introduction The UNI-T Model UT713 Thermocouple Calibrator is a precise source and measurement tool for calibrating thermocouple instruments. The calibrator sources or measures mV or 8 different type of TC in units of , . But the Meter cannot uses as an output source and measurement.
Model UT713: OPERATING MANUAL Safety Information This Meter complies with the standard CE EN61326. Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired. Warning To avoid possible electric shock or personal injury: l Before using the calibrator inspect the case. Do not use the calibrator if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic.
Model UT713: OPERATING MANUAL l Replace the battery as soon as the battery indicator appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury. l The internal circuit of the calibrator shall not be altered at will to avoid damage of the calibrator and any accident l Do not use or store the calibrator in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field.
Model UT713: OPERATING MANUAL Turning the Calibrator On Press to turn the calibrator on and off Output Show TC type or voltage range selected Press to toggle or Press so that OUTPUT is on the display Press TYPE to select TC type or DC mV Press to turn the Meter on and off Press and to increment or decrement a digit. Press and to step up/down from 0 to 9 .
Model UT713: OPERATING MANUAL Input Show TC type or voltage range elected Display the selected or Press so that INPUT is on the display Press to toggle or Press to turn the Meter on and off Press Type to select TC type or DC mV Input Terminals Figure 2 7
Model UT713: OPERATING MANUAL 1. Press the pushbutton to turn the Calibrator on. 2. Press OUTPUT to display OUTPUT if the Calibrator is at input mode which displays INPUT. 3. Press to select the requested TC type or DC mV range. 4. Press or to increment or decrement a digit. 5. Press or to step up or down from 0 to 9. Hold down to scroll faster. 6. Insert the test leads to the OUPUT terminals as figure 3. 7.
Model UT713: OPERATING MANUAL Measurement equipment Figure 3 9
Model UT713: OPERATING MANUAL 1. Press the pushbutton to turn the Calibrator on. 2. Press INPUT to display INPUT if the Calibrator is at output mode which displays OUTPUT. 3. Press to select the requested TC type or DC mV range. 4. Insert the test leads to the INPUT terminals as figure 4. 5. Connect the other end of the test leads to the TC simulate equipment or the DC mV source equipment. Note: Measuring TC must select the corresponding TC type.
Model UT713: OPERATING MANUAL Output equipment Figure 4 11
Model UT713: OPERATING MANUAL Follow the below procedure to start the cold terminal temperature auto compensation functions: 1. Press the pushbutton to turn the Calibrator on. 2. Press INPUT to display enter INPUT mode 3. Press to select J type. 4. Press and hold and together to display 0 or 1 (display 0 is cold terminal temperature auto compensation, display 1 is the remove cold terminal temperature auto compensation), release the two buttons. 5. Press and to adjust the digit to 0, press to confirm. 6.
Model UT713: OPERATING MANUAL Note: Because the temperature transducer aluminum plate is put at the back of the Calibrator, try to open the tilt stand if possible when operating the Calibrator. Don’t touch the aluminum plate when operating the Calibrator to ensure the accurate of compensation temperature. If the operating temperature is changed, it is necessary to wait for around 10 minutes to stable the internal transducer before operating the Calibrator.
Model UT713: OPERATING MANUAL Explanation of International Symbols The following symbols are used on the calibrator or in this operating manual. The table below explains their meaning. International Symbols Symbol 14 Meaning Grounding Warning.
Model UT713: OPERATING MANUAL Maintenance Warning Make sure the calibrator is off and the test leads are removed from the input terminals and the circuit under test before opening the calibrator’s case. For maintenance procedures not described in this sheet, contact your dealer. l Do not store the Meter in an environment of high temperature, humidity and strong magnetic field. l Check the battery and test leads. Replace as necessary.
Model UT713: OPERATING MANUAL Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents. l When the symbol appears on the display, replace the battery with a 9V alkaline battery (1604A or 6LF22) l Remove the battery from the battery door and turn the calibrator off when it is not using for a long time.
Model UT713: OPERATING MANUAL Warning To avoid personal injury or damage to the calibrator, use only a 125mA 250V fuse for F1. In Input mode, the calibrator and TC are reliably connected. When the display continuisly shows OL, the fuses may be blown.
Model UT713: OPERATING MANUAL Replace the fuses using the following procedure: 1. Remove the test leads from the calibrator terminals and turn the calibrator off. 2. Remove the screw on the battery door, then remove the battery door and take out the battery. 3. Remove the three screws from the case bottom and turn the case over. 4. Gently remove the fuse from its mounting bracket. 5. Replace the blown fuse with a 125mA 250V fuse. 6. Fit the top and bottom covers together. 7. Reinstall the three screws. 8.
Model UT713: OPERATING MANUAL Specifications Specifications are based on a one year calibration cycle and apply for ambient temperature from +18 to +28 unless stated otherwise. Thermocouple Type J K T E N Temperature Ranges -200.0~1200.0 -328.0~2192.0 -200.0~1370.0 -328.0~2498.0 -200.0~400.0 -328.0~752.0 -200.0~950.0 -328.0~1742.0 -200.0~1300.0 -328.0~2372.0 Display Resolution 0.1 or 0.1 Accuracy 0.04% ( -100 0.04% ( -148 0.04% ( -100 0.04% ( -148 1.5 ) 2.7 ) 1 ) 1.
Model UT713: OPERATING MANUAL Thermocouple Type Temperature Ranges Display Resolution -20~1750 R -4~3182 -4~3182 600~1800 B 20 1112~3272 0.04% 100 0.04% ( 212 0.04% ( 100 0.04% ( 212 ( -20~1750 S Accuracy 1 or 1 3 ) 5.4 ) 2 ) 3.
Model UT713: OPERATING MANUAL Thermocouple Type Standard Scales J, K, T, E, N, R, S, B NIST-175 ITS-90 Mode Range Display Resolution mV -10.00~110.00mV 10 V -110.0~1100.0mV 0.1mV Accuracy 0.
Model UT713: OPERATING MANUAL Resolution: TC: 0.1 or 0.1 (J, K, T, E, N); 1 or 1 (R, S, B). DC mV: 10 V (110mV range); 0.1mV (1100mV range) Cold Junction Error: 0.5 (-0~50 ) Maximum voltage applied between any terminal and earth ground or between any two terminals: 30V Storage temperature: -10 to 55 Operating temperature: -0 to 50 Operating altitude: 3000 meters maximum Temperature coefficient:0.005% x specified range per for temperature ranges -0 to 18 and 28 to 50 .
Model UT713: OPERATING MANUAL * END * This operating manual is subject to change without notice.
Model UT713: OPERATING MANUAL Copyright 2007 Uni-Trend Group Limited. All rights reserved. Manufacturer: Uni-Trend Technology (Dongguan) Limited Dong Fang Da Dao Bei Shan Dong Fang Industrial Development District Hu Men Town, Dongguan City Guang Dong Province China Postal Code: 523 925 Headquarters: Uni-Trend Group Limited Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.