Datasheet

Specifications Range
Model UT61A UT61B
400mV/4V/40V/400V/1000V ±(0.8%+1)
40mV/400mV/4V/40V/400V/1000V ±(0.5%+1)
400mV/4V/40V/400V/750V ±(1%+3)
40mV/400mV/4V/40V/400V/750V ±(1%+3)
DC Current (A) 400μA/4000μA/40mA/400mA/4A/10A ±(1%+2) ±(1%+2)
AC Current (A) 400μA/4000μA/40mA/400mA/4A/10A ±(1.2%+5) ±(1.2%+5)
Resistance (Ω) 400Ω/4KΩ/40KΩ/400KΩ/4MΩ/40MΩ ±(1%+2) ±(1%+2)
Capacitance (F) 40nF/400nF/4μF/40μF/400μF/4000μF ±(3%+5) ±(3%+5)
Frequency (Hz) 10Hz-10MHz ±(0.1%+4) ±(0.1%+4)
Temperature (°C)
-40 °C
1000°C
±(1.2%+3)
Temperature (°F)
-40°F1832°F
±(1.2%+4)
Display Count 4000 4000
Auto Range
Manual Range
Bandwidth (Hz 45Hz 400Hz
Duty Cycle
0.1% 99.9%
Diode
Transistor
NCN Function
Auto Power Off Around 15 Minutes
Continuity Buzzer
Low Battery Indication
7.5V
Data Hold
Relative Mode
MAX/MIN
RS-232
LCD Backlight
Analogue Bar Graph
41 41
Input Protection
Input Impedance for DCV Around 10MΩ
Power
LCD Size
Product Color
Product Net Weight
Product Size
Standard Individual Packing
Standard Quantity Per Carton
Standard Carton Measurement
Modern Digital Multimeters
UT61A UT61B
Introduction
UT61 Series
Safety Rating
Certifications
GS
CE
EN:61010-1
EN:61326-1
Best Accuracy
CAT III 1000V
CAT IV 600V
180mm X 87mm X 47 mm
DC Voltage (V)
Features
General Characteristic
AC Voltage (V)
UT61A-Battery, Test Lead,Muli-Purpose Socket
9V Battery
6F22)
65mm X 43 mm
Red and Grey
465mm X 335mm X 345mm
370g
UT61B-Battery, Test Lead,Muli-Purpose Socket,RS-232 Interface
Cable,Software,Point Contact Temperature Probe
Standard Accessories
Gift Box, English Manual
20 PCs