Model UT60H: OPERATING MANUAL Table of Contents Page Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons Display Symbols Measurement Operation A. DC Voltage Measurement B. AC Voltage Measurement C. DC Current Measurement D. AC Current Measurement E. Measuring Resistance F. Testing for Continuity G. Testing Diodes H. Capacitance Measurement I. Frequency Measurement J.
Model UT60H: OPERATING MANUAL Title Page Accuracy Specifications A. DC Voltage B. AC Voltage C. DC Current D. AC Current E. Resistance F. Continuity Test G. Diode Test H. Capacitance I. Frequency J. Temperature K. 4~20mA Loop Maintenance A. General Service B. Replacing the Fuses C.
Model UT60H: OPERATING MANUAL Overview This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly. Warning To avoid electric shock or personal injury, read the “Safety Information” and “Rules for Safe Operation” carefully before using the Meter.
Model UT60H: OPERATING MANUAL Unpacking Inspection Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part: Qty Item Description 1 piece 1 English Operating Manual 2 Test Lead 1 pair K-Type (nickel chromium ~ nickel silicon) 1 piece 3 Point Contact Temperature Probe 4 Multi-Purpose Socket 1 piece 5 Alligator Clip 1 piece 6 0.
Model UT60H: OPERATING MANUAL 5
Model UT60H: OPERATING MANUAL Rules For Safe Operation (1) Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules: l Before using the Meter inspect the case. Donot use the Meter if it is damaged or the case (orpart of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. l Inspect the test leads for damaged insulation or exposed metal.
Model UT60H: OPERATING MANUAL Rules For Safe Operation (2) l l l l l l l l l l Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads with identical model number or electrical specifications before using the Meter.
Model UT60H: OPERATING MANUAL Rules For Safe Operation (3) l l l l l l l l Before measuring current, check the Meter’s fuses and turn off power to the circuit before connecting the Meter to the circuit. Replace the battery as soon as the battery indicator appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury. When servicing the Meter, use only the same model number or identical electrical specifications replacement parts.
Model UT60H: OPERATING MANUAL International Electrical Symbols AC or DC. Grounding. Double Insulated. Warning. Refer to the Operating Manual.Deficiency of Built-In Battery. Conforms to Standards of European Union.
Model UT60H: OPERATING MANUAL The Meter Structure (see figure 1) 1. 2. 3. 4. EF Sensor (front part of the housing) LCD Display Functional Buttons Rotary Switch 5.
Model UT60H: OPERATING MANUAL Rotary Switch Below table indicated for information about the rotary switch positions. Rotary Switch Function Position AC or DC voltage measurement V Hz o C % EF µA : Resistance measurement : Diode test : Continuity test Capacitance measurement Frequency measurement Temperature measurement (UT60G only) 4~20mA percent measurement Sensor Test AC 50-400V AC signal AC or DC current measurement (0.1µA ~4000µA) mA AC or DC current measurement (0.01mA ~ 400.
Functional Buttons
4 5 6 7 8 9 10 11 3 2 1 15 14 13 12
Model UT60H: OPERATING MANUAL Measurement Operation(1) A. DC Voltage Measurement (see figure 3) ( figure 3) Warning To avoid harms to you or damages to the Meter from eletric shock, please do not attempt to measure voltages higher than 1000V, although readings may be obtained. The DC Voltage ranges are: 4V, 40V, 400V and 1000V; To measure DC voltage, connect the Meter as follows: 1. Insert the red test lead into the V terminal and the black test lead into the COM terminal. 2.
Model UT60H: OPERATING MANUAL Measurement Operation(3) C. DC Current Measurement (see figure 4) ( figure 4) Warning Never attempt an in-circuit current measurement where the open circuit voltage between terminals and ground is greater than 60V DC or 30V rms. If the fuse burns out during measurement, the Meter may be damaged or the operator himself may be hurt. Use proper terminals, function, and range for the measurement.
Model UT60H: OPERATING MANUAL Measurement Operation(4) Note l If the value to be measured is unknown, use the maximum measurement position and reduce the range step by step until a satisfactory reading is obtained. l When the measured current is 5A, continuous measurement is allowed. l When the measured current is between 5A-10A, continuous measurement 10 seconds and interval more than 15 minutes.
Model UT60H: OPERATING MANUAL Measurement Operation(5) Note l If the value to be measured is unknown, use the maximum measurement position and reduce the range step by step until a satisfactory reading is obtained. l When the measured current is 5A, continuous measurement is allowed. l When the measured current is between 5A-10A, continuous measurement 10 seconds and interval more than 15 minutes.
Model UT60H: OPERATING MANUAL Measurement Operation(6) E. Measuring Resistance (see figure 5) (figure 5) Warning To avoid harms to you, please do not attempt to input voltage higher than 60V DC or 30V rms AC. To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance. The resistance ranges are:400Ω, 4kΩ, 40kΩ,400kΩ, 4MΩand 40MΩ. To measure resistance, connect the Meter as follows: 1.
Model UT60H: OPERATING MANUAL Measurement Operation(7) 21
Model UT60H: OPERATING MANUAL Measurement Operation(8) F. Testing for Continuity (see figure 6) (figure 6) Warning To avoid harms to you, please do not attempt to input voltage higher than 60V DC or 30V rms AC. To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring continuity. Note l Open circuit voltage around –1.2V and range is 400Ω measurement range.
Model UT60H: OPERATING MANUAL Measurement Operation(9) G. Testing Diodes (see figure 7) ( figure 7) Warning To avoid harms to you, please do not attempt to input voltages higher than 60V DC or 30V rms AC. To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before testing diodes. Use the diode test to check diodes, transistors, and other semiconductor devices.
Model UT60H: OPERATING MANUAL Measurement Operation(10) Note l In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8V; however, the reverse voltage drop reading can vary depending on the resistance of other pathways between the probe tips. l Connect the test leads to the proper terminals as said above to avoid error display. l The LCD will display OL indicating either open circuit or wrong polarity connection.
Model UT60H: OPERATING MANUAL Measurement Operation(11) Ω Note l The LCD displays OL indicating the tested capacitor is shorted or it exceeds the maximum range. l Analogue bar graph will be disabled when measuring capacitance. l It is normal to take a while when measuring high capacitor 400 F. The analogue bar graph at this time will show the time left before finishing the measurement.
Model UT60H: OPERATING MANUAL Measurement Operation(12) I. Frequency Measurement (see figure 9) ( figure 9) Warning To avoid harms to you, please do not attempt to input voltage higher than 30V rms. The measurement ranges are:40Hz, 400Hz, 4kHz, 40kHz, 400kHz, 4MHz, 40MHz and 400MHz. To measure frequency, connect the Meter as follows: 1. 2. 3. Insert the red test lead into the Hz terminal and the black test lead into the COM terminal. Set the rotary switch to Hz.
Model UT60H: OPERATING MANUAL Measurement Operation(13) J. Temperature Measurement (see figure10) ( figure 10) Warning To avoid harms to you, please do not attempt to input voltages higher than 60V DC or 30V rms AC. o o The temperature measurement range is -40 C~400 C. To measure temperature, connect the Meter as follows: o 1. Set the rotary switch to C, the LCD displays room temperature. 2. Insert the multi-purpose socket into the Meter as the figure 10. 3.
Model UT60H: OPERATING MANUAL Measurement Operation(14) K. EF Function (see figure 11) ( figure 11) The EF function can detect AC 50~400V. To use EF function, connect the Meter as follows: o 1. Set the rotary switch to C/EF and press SELECT to select EF measurement mode. 2. Place the housing front part near the object being measured. 3. The LCD displays “EF”. The Analogue bar graph indicates the AC signal between 50~400V.
Model UT60H: OPERATING MANUAL Sleep Mode To preserve battery life, the Meter automatically turns off if you do not turn the rotary switch or press any button for around 30 minutes. The Meter enters into sleep mode The Meter can be activated by turning the rotary switch or pressing any button except POWER and Low imp. 200KΩ button. To disable the Sleep Mode function, press MAX MIN or REL button while turning on the Meter.
Model UT60H: OPERATING MANUAL The Use of Relative Value Mode The REL mode subtracts a stored value from the present measurement value and displays the result. For instance, if the stored value is 20.0V and the present measurement value is 22.0V, the reading would be 2.0V. If a new measurement value is equal to the stored value then display 0.0V.
Model UT60H: OPERATING MANUAL Low Impedance 200kΩ Button Warning This button can only applies on testing voltages lower than 250Vmax and press down the button not more than 3 seconds. When the Meter is under V mesaurement ranges, press Low imp. 200 kΩ button to switch over the input impedance from 10MΩ to around 200kΩ.
Model UT60H: OPERATING MANUAL General Specifications 32
Model UT60H: OPERATING MANUAL A. DC Voltage Range Resolution 4V 40V 400V 1000V 0.0001V 0.001V 0.01V 0.1V Accuracy (0.1%+5) Overload Protection 1000V (0.2%+5) Remarks: l Input impedance: Approx.10MΩ . B. AC Voltage Range Resolution 4V 40V 400V 750V 0.001V 0.01V 0.1V 1V Accuracy (1%+4) Overload Protection 1000V (1.5%+4) Remarks: l Input impedance: Approx 10MΩ . l Display: a) True rms value (suitable on range’s 10%~100% ) b) AC wave peak factor:3.0 (1.
x x
“ ”
Model UT60H: OPERATING MANUAL MAINTENANCE(1) This section provides basic maintenance information including battery and fuse replacement instruction. Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. A. General Service l l l l l Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
Model UT60H: OPERATING MANUAL MAINTENANCE(2) Warning To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure. To replace the Meter’s fuse: l Press the POWER to turn the Meter off and remove all connections from the terminals. l Remove the fixed rubber feet and 5 screws from the case bottom. l Remove the fuse by gently prying one end loose, then take out the fuse from its bracket.
Model UT60H: OPERATING MANUAL MAINTENANCE(3) Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ” appears. Make sure the test leads are disconnected from the circuit being tested before opening the case bottom. To replace the battery: l Press the POWER to turn the Meter off and remove all connections from the terminals.
Model UT60H: OPERATING MANUAL 41
Model UT60H: OPERATING MANUAL Copyright 2003 Uni-Trend International Limited. All rights reserved. Manufacturer: UNI-TREND TECHNOLOGY(DONG GUAN)LIMITED Address: Dong Fang Da Dao, Bei Shan Dong Fang Industrial Development District, Hu Men Town, Dong Guan City, Guang Dong Province, China Headquarters: Uni-Trend International Limited Address: Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.