Specifications
Maintenance (4)
D. Replacing the Fuses (see figure 11)
(figure 11)
f
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
To replace the Meter's fuse:
1. Turn the Meter to OFF position and remove all
connections from the terminals.
2. Remove the screw from the battery compartment,
and separate the battery compartment from the case
bottom.
3. Remove the 2 rubber feet and 2 screws from the
case bottom, and separate the case top from the
case bottom.
4. Remove the fuse by gently prying one end loose,
then take out the fuse from its bracket.
5. Install ONLY replacement fuses with the identical
type and specification as follows and make sure the
fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 250V, fast type Glass fuse,
5x20mm.
Fuse 2: 10A, 250V, fast type Glass fuse,
5x20 mm.
Model UT60D: OPERATING MANUAL
37