Measurement Operation (1)
A. AC & DC Voltage Measurement
Warning
To avoid harms to you or damages to the Meter from
electric shock, please do not attempt to measure
voltages higher thanDC1000V / AC750V rms, although
readings may be obtained.
AC Voltage Measurement (see figure 3)
(figure 3)
The AC voltage ranges are: 3.400V, 34.00V, 340.0V and
750.0V. To measure AC Voltage, connect the Meter as
follows:
1. Insert the red test lead into the HzV terminal and
the black test lead into the COM terminal.
2. Set the rotary switch to V
Hz and press BLUE
button to select AC measurement mode.
3. Connect the test leads across with the object being
measured.
The measured value shows on the display, which is
effective value of sine wave (mean value response).
Note
l In each range, the Meter has an input impedance
of 10M . This loading effect can cause
measurement errors in high impedance circuits. If
the circuit impedance is less than or equal to 10k ,
the error is negligible (0.1% or less).