Specifications
Digital Multimeter
Page <24> 20/09/10 V1.1
http://www.farnell.com
http://www.newark.com
http://www.cpc.co.uk
Accuracy Specifications (3)
Overload Protection:
Glass fuse 0.5A, 250V, fast type, 5 x 20mm.
H. Temperature (Model UT60C)
Range Resolution Accuracy
-40°C to 1000°C 1°C
-40°C to 0°C ±(3% + 4)
0°C to 400°C ±(1% + 3)
400°C to 1000°C ±(2% + 10)
I. DC Current
Range Resolution Accuracy Overload Protection
400µA 0.1µA
±(1% + 2)
0.5A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
4000µA 1µA
40mA 0.01mA
±(1.2% + 3)
400mA 0.1mA
4A 0.001A
±(1.5% + 5)
10A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
10A 0.01A
Remarks:
• 4A and 10A Range:
For continuous measurement
≤≤
10 seconds and interval not less than 15 minutes.
Accuracy Specifications (4)
J. AC Current
Range Resolution Accuracy Overload Protection
400µA 0.1µA
±(1.5% + 5)
0.5A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
4000µA 1µA
40mA 0.01mA
±(2% + 5)
400mA 0.1mA
4A 0.001A
±(2.5% + 5)
10A, 250V,
fast type Glass fuse,
Ø5 x 20 mm
10A 0.01A
Remarks:
Frequency response 40Hz to 400Hz.
UT60C displays effective value of since wave (mean value response) displays true rms
value.
• 4A and 10A Range:
For continuous measurement
≤≤
10 seconds and interval not less than 15 minutes.