UT581/582 Digital RCD (ELCB)Tester Instruction Manual
UT581/582 INSTRUCTION MANUAL CONTENTS Chapter Page 1.Safety Instruction 2 2.Characteristics 6 3.General Specifications 8 4.Product Outlook & Accessories 12 5.Measurement Operation 14 6.
UT581/582 INSTRUCTION MANUAL 1. Safety Instructions This operation manual includes the user guidance and safety instruction when using the tester, please read it before using. Caution ● Before using the tester, please read and understand the operating manual including the content. ● Keep the operating manual properly, and let it easy to get it for reference during the process of testing. ● When using the tester, user must follow the testing procedure as mentioned in the operating manual.
UT581/582 INSTRUCTION MANUAL Safety sign“ ”has three meaning in this manual, user has to pay the attention to this sign“ ”for operation. Danger---identifies conditions and actions that most likely pose hazard(s) or die. Warning---identifies conditions and actions that will pose hazard(s) or die. Caution---identifies conditions and actions that will pose hazard(s) or damage the tester. Danger ● This test instrument unit is suitable for using under single phase 230V+10%-15%.
UT581/582 INSTRUCTION MANUAL ● Do not use your finger to touch on any testing cable during the testing stage. ● Put off the test leads after testing complete. P.S. test leads can not always plugs on the test instrument. Warning ● Do not open or disassemble the tester un-intentionally during the measurement, it has a high voltage inside and cause a danger for the user. If it needs service repairing, please contact our after-sales services or our agents.
UT581/582 INSTRUCTION MANUAL Caution ● Ensure the test lead probe (need to be provided by UNI-TREND Certified One) insert into the corresponding port in order to provide a safe before measurement. It is not encourage to use third parties test lead probe on our tester. ● Do not expose the tester in extreme temperature and wet environment. ● Soft cloth and mild detergent should be used to clean the surface of the tester. No abrasive and solvent should be used when servicing.
UT581/582 INSTRUCTION MANUAL 2. Characteristics 2.0 This tester is suitable for using under single phase 230 V / 50 Hz. (Operational Voltage Scope: 195 – 253 V) 2.1 It uses micro-controller with high accuracy, reliability and satiability. 2.2 Connectivity Checking: 3-LED Lamp signal is correct or not . Correct connectivity: P-E & P-N display two green light signal on but red light is off, If not, you may incorrect to connect it. 2.
UT581/582 INSTRUCTION MANUAL in sync(UT582 only). 2.8 Power Off Buzzer: Buzzer alarm after turn off 3 minutes. 2.9 Environmental Protection Energy Conservation: This tester can not use Battery operation, it can use cable to charge on under(under A/C 220V/50Hz only). 2.10 Fuse: Fused safety protection directly. 2.11 Double insulation or reinforced insulation safety manufacturing.
UT581/582 INSTRUCTION MANUAL 3. General Specifications 3.1 Measuring range & Tolerance.
UT581/582 INSTRUCTION MANUAL UT582: Functions Voltage(AC) X 1/2 X1 X5 AUTO RAMP TEST Trip Current Setting (I△n) Trip Time (MAX) 10 / 20 / 30 /100 / 300 / 500mA 230V (Tolerance: 10 / 20 / 30 /100 / -15%~+10%) 300 mA Frequency: 500mA 50Hz 10 / 20 / 30mA Accuracy Trip Current Trip Time 1000mS Tolerance: -10%~0% 1000mS ±0.
UT581/582 INSTRUCTION MANUAL 3.2 Measuring Range (Functions) ×1/2 as disconnecting measurement, test RCD sensitive. ×1 as connecting measurement of the response time. ×5 as fast connecting measurement of I△n×5 time. 250mA as connecting measurement of the response time. (UT581 only) AUTO RAMP TEST as connecting measurement of the size of the current. (UT582 only) 3.3 Application Standard: IEC 61010-1 CAT III 300V Polluting Grade: Grade II IEC 61557-1,5 IEC 61010-2-31 3.
UT581/582 INSTRUCTION MANUAL 3.5 Working Environment: Temperature: 0℃~40℃ Relative Humidity: ≤80%RH Elevation: ≤2000 meter 3.6 Storage Condition: Temperature: -20℃~60℃ Relative Humidity: ≤75%RH 3.7 Product Size: 160mmx70.5mmx100mm 3.8 Product Net Weight: About 500g 3.9 Standard Accessories: Test Leads(1.
UT581/582 INSTRUCTION MANUAL 4.
UT581/582 INSTRUCTION MANUAL Red (Live) Green ( Earth ) Blue ( Neutral ) Diagram 2 ① ② ③ ④ LCD Display PHASE Button I△n Button TEST Button ⑤ ⑥ ⑦ ⑧ Connecting Lead Testing LED Signal Lamp Rotary Mode Switch Test Terminal Test Lead 13
UT581/582 INSTRUCTION MANUAL 5. Measurements 5.1 Test Lead Connection Test terminal is / are for connecting the correct lead of the tested installation: L to L(Live); N to N(Neutral); E to E(Earthing). 5.2 Test Lead Checking After connected the test leads with the testers, please plug the power cable on (220V/50Hz current socket), so you can 3-LED lamp whether it is correct. If P-E & P-N are displayed green signal light, but the red color LED do not light up. It is in good normal condition.
UT581/582 INSTRUCTION MANUAL Danger ● If the connecting measurement is incorrect, please don’t proceed the next measurement (press TEST button). Otherwise, it causes measuring result incorrect or in other dangerous. 5.3 Press I△n button switch, it causes RCD breakers for the nominal residual currents (I△n) as the previous trip current in constant Below LCD display the default connecting current value. Default Value: I△n 30mA 0°/180° 0° 5.4 Measuring Methods 5.4.
UT581/582 INSTRUCTION MANUAL • Rapid Connecting Test ×5 (10,20,30mA only): Maximum Measuring Time 1000ms • Rapid Connecting Test 250 mA: Maximum Measuring Time1000ms(UT581 only) • AUTO RAMP TEST Auto Ramp Test(UT582 only). 20%~110% Default nominal residual currents(I△n). Maximum measuring time 300ms×10. 5.4.2 Press TEST(Measure)button • Disconnecting Test Breaker should not be connected. •Connecting Test Breaker should be connected. •Fast Connecting Test Breaker should be connected.
UT581/582 INSTRUCTION MANUAL 5.4.3 Press PHASE (0°/180°) button, phase switch as repeated Step (5.4.2 ) to get the fast connecting time. 5.4.4 Need to phase switch and repeated the Step (5.4.2). Warning ● Do not use the tester if it is damaged or metal part is exposed. ● During the long hours and continually use, the tester may arise in more heating and cause to the unit damage or / and other danger arises, so we do not encourge to use in long hours and it is used on sample random testing purpose only.
UT581/582 INSTRUCTION MANUAL 6. Maintenance & Repair 6.1 Soft cloth and mild detergent should be used to clean the surface of the tester because solvent will corrosive the display and avoid moisture. 6.2 Repair Contact our after-sales service department or agent when the following thing happens: A.The tester case is being damage or broken. B.LCD display is in abnormal. C.Unreasonable deviation when in normal use. D.Buttons do not function properly and confusion. E.Sound during the testing.
UT581/582 INSTRUCTION MANUAL 19