Specifications
7
B-5 Resistance
200Ω
2KΩ
20KΩ
200KΩ
2MΩ
20MΩ
0.01Ω
0.1Ω
1Ω
10Ω
100Ω
1KΩ
Range Resolution Accuracy
(0.1% +2)
(0.1% +5)
(0.3% +10)
Overload protection: All ranges are 230VDC
or AC RMS.
Voltaeg at Open Circuit: About 3V
Caution: At 200Ω range, short the test lead
first to obtain the resistance value of the test
leads. Deduct the resistance value from the
measured reading to obtain the correct value
of the object being measured .
B-6 Capacitance
20nF
200nF
2µF
20µF
1pF
10pF
0.1nF
1nf
Range Resolution Accuracy
(2.0% +10)
(2.5% +30)
Testing Signals: About 400Hz, 40mVrms.
Model UT57: OPERATING MANUAL