Operating instructions

9. Communication with external data acquisition system (option)
page 94
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
<STX>;P;19.2;43850;<ETX> temp. of calibration cell = 19.2 °C
<STX>;P;K;51699;<ETX>
<STX>;P;E;43511;<ETX> E = single (manual) calibration
<STX>;P;I;43506;<ETX>
<STX>;P;U;27130;<ETX> U = µl (when manual calibration has been chosen)
<STX>;P;P;14841;<ETX> P = pg (when manual calibration has been chosen)
<STX>;P;4;63955;<ETX> calibration. interval = 4 hours (when Auto. Calibration has
been chosen)
<STX>;P;1130;44505;<ETX> start time of calibration. = 11:30 (when auto calibration. Has
been chosen)
<STX>;P;D;14838;<ETX>
<STX>;P;280204;1250;56724;<ETX> date = 28.02.04
time = 12:50 ( of last calibration)
<STX>;P;C;2548;<ETX>
<STX>;P;1.23456789;2.34567;3.45678;10851;<ETX> calibr. Data (equation; const;
correlation coefficient)
<STX>;P;V;39418;<ETX>
<STX>;P;1.2;280204;1040;708;<ETX> last concentration = 1.2, date = 28.02.04,
time= 10:40
System control command:
Inquiry: <STX>;C;S;<CRC>;<ETX> C = control
S = start span-check
Answer: <STX>;C;S;<CRC>;<ETX> span-check will be started
Example: <STX>;C;S;19965;<ETX> 19965 = CRC16
<STX>;C;S;19965;<ETX> span-check will be started
Inquiry: <STX>;C;S;V;<CRC>;<ETX> C = control
S = span-check
V = value (= read span-check value)
Answer: <STX>;C;V;conc;<CRC>;<ETX> V = value
conc = meas. concentration in ng/m³
Example: <STX>;C;S;V;55038;<ETX> 55038 = CRC16
<STX>;C;V;1.3;14490;<ETX> 14490 = CRC16
1.3 = concentration
Inquiry: <STX>;C;Z;<CRC>;<ETX> C = control
Z = start zero-check (sample volume= 1litre)
Answer: <STX>;C;Z;<CRC>;<ETX> zero-check will be started
Example: <STX>;C;Z;7675;<ETX> 7675 = CRC16
<STX>;C;Z;7675;<ETX> zero-check will be started