Operating instructions

7. Technical Information
page 78
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
7. Technical Information
7.1 Technical specifications
Instrument type
UT-3000
Manufacturer
Mercury Instruments GmbH, Liebigstr. 5,
85757 Karlsfeld, Germany
Measuring component
Total Gaseous Mercury (TGM)
Measurement principle
Atomic absorption, cold vapor method,
Preconcentration with GoldTrap
Measurement wavelength
253.65 nm
UV-source
electrodeless mercury discharge lamp
Method of stabilising
optically with reference beam; thermally
Optical cell
entirely of fused silica Suprasil, l= 230 mm
Cell temperature
ca. 70 °C, heated
Measuring range a) lowest range
b) medium range
c) screening mode
0.1 ng/m³-1000 ng/m³ at 10 l sample volume
1 ng/m³-3000 ng/m³ at 1 l sample volume
10 ng/m³ - 30000 ng/m³ at 0.1 l sample
volume
Detection limit according DIN 32645
1 pg absolutely or 0.1 ng/m³
Linearity over the 0.2 to 15000 ng/m³ range 1.5 % or better
Reproducibility 5 % or better at 14 pg Hg
1 % or better at 1500 pg Hg
Sampling flow 40 l/h
Goldtrap heating flow
10 l/h
Digital output signal
Serial (USB/RS 232) for PC connection,
Data logger capacity 5000 data sets
Electrical power consumption
Photometer: max. ca. 250 VA
Calibration Unit: ca. 70 VA
Power supply
230 VAC / 50 Hz ; 115 VAC / 60 Hz
(option) ; 12 V DC (option)
Dimensions
45 x 15 x 35 cm (WxHxD) without filter
Weight
ca. 12 kg
7.2 Influencing effects
Operating temperature
0 °C ... 40 °C
Operating humidity
max. ca. 90 % R.H., no condensing
humidity
CE approval
according to 89/336 and 73/23 EEC