Operating instructions

5. Service-Menu
page 54
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
5.11 Special Service Codes
Following codes are available:
Code:
Code:Code:
Code:
Function:
Function:Function:
Function:
Recommended Setting
Recommended SettingRecommended Setting
Recommended Setting
321
Standard Service Menu, see beginning of this chapter n.a.
191
Standard / Tedlar Bag Mode Standard (Tedlar Bag Mode is
an option for special
applications)
222
Offset T Cal.Cell / Zero Offset / Heating Flow /
Heating Flow Regulate
0.00 °C / 0.0000 / 10 l/h /ON
333
Evaluation Graph on/off ON
753 Flow Calibration (10 l/min – 3 points; 40 l/min – 3
points)
Only to be used for service !!!
(service tools required !!!)
777
Calib. Syringe on/off As required by user
881
Absorbance correction function (A-factor) 1.000
987 allows to actuate valves For service only
999
Automatic Calibration / Span Calibration As required
Remarks:
191: Tedlar-Bag-Mode is only used for instruments optionally equipped for this method to
determine Hg in natural gas or other gas samples.
222: Zero offset- for correction of the analytical zero value, this value is subtracted from the
measurement peaks.
Temp offset- for correction of the offset of an automatic or manual calibration device.
Heating Flow: allows to set heating flow, shall not be changed from standard setting except
for special applications (high measuring range).
Flow Regulate: For service purpose only, shall not be changed from standard setting.
If Heating flow regulate= ON“ then the pump is controlled by the mass flow meter signal
within +/- 2 l/h of the set value.
If Heating flow regulate= OFF“, then the pump is fed with a constant voltage (7.5 V) and the
flow may vary within a range of +/- 3 l/h.
If the Flow is out of range, a Repeat Zero will be performed up to 10 times, after this the system
will be halted and an error message will be displayed.
333: Shows the peak during System Evaluation when on; a horizontal line is displayed when off.
753: For calibration service of flow meter only!
777: Only on if an Automatic Calibration Unit is installed. Has to be on if an automatic
calibration should be performed.
881: For service only!
987: For service only!; allows to actuate valves and status contacts.
999: Mode of automatic calibration can be selected: multiple point calibration or one point span
calibration.