Operating instructions

5. Service-Menu
page 52
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
5.6.2 Set Measuring Calibration
By pressing F2 – Set Meas-Calibration you can enter manually the values of the measuring
calibration, in case they are deleted.
5.6.3 Set T-Cal. Cell Calibration
By pressing F3 – Set T-Cal.Cell Calibration you can enter manually the values for the
calibration of the temperature sensor of the calibration device (either manual or automatic
calibrator).
5.7 Trap Counter
This window displays the total number of heating cycles of the GoldTrap ever carried out.
5.8 Set RS232 Mode
As a standard the RS232 output delivers measurement data sequentially as ASCII characters.
Optionally data can be output following a protocol for data acquisition systems from ISEO
(France).
5.9 Set Zero Duration
The instrument performs an automatic zero adjustment before each trap heating step (see also
chapter 3.5). In this menu the duration of the zeroing can be set to a value between 11 seconds
and 250 seconds. Standard setting is 11 seconds. This parameter has to be increased if during
operation of the system with real sample it is found that the reading towards the end of zeroing is
ESC <<< Back
Service Mode
F1
Set gradient
F2
Set constant
Y = .......X +/- 0.......