Operating instructions

5. Service-Menu
page 51
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
5.5 Change Syringe (for Automatic Calibration Option only)
This function is only available when the Automatic Calibration function has been activated
(Service code 777). It is necessary for change of the syringe inside the Automatic Calibration
Unit. The command Change Syringe moves the piston of the syringe to a 50 % position which is
necessary to remove the syringe. The Reset command pushes the piston back again into the
syringe.
5.6 Set Calibration Data
In this window calibration parameters for the built in mass flow sensor and the parameters of the
analytical calibration can be set. This is useful if these values are lost due to service work.
5.6.1 Set Flow Calibration
By pressing F1 – Set Flow-Calibration you can enter manually the calibration data of the
internal mass flow controller, in case they are deleted (for example if the microprocessor-battery
is empty). You will find those data on the test-certificate of your UT-3000, which was in the
delivery amount.
ESC <<< Back
Service Mode
Set Calibration Data
F1
Set Flow-Calibration
F2
Set Meas-Calibration
F2
Set T-Cal. Cell Calib.