Operating instructions
4. Calibration
page 46
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
4.5.5 Manual Start of the Automatic Calibration
F3- Start automatic Calibration is selected in the Calibration Mode menu. Now an automatic
calibration is immediately started using the concentrations entered under F1-Cal. Concentration.
The following screens can be noticed consecutively:
After that the standard cleaning routine of the GoldTrap is done:
Trap Cleaning / Trap Cooling / Zeroing / Trap Cleaning / Trap Cooling / Zeroing /
Trap Cleaning / Trap Cooling
Now the routine for each calibration point is starting and repeated as often as your amount of
different concentrations.