Operating instructions
4. Calibration
page 39
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
ESC <<< Back
F1
F2
Enter pg
Enter µl
Manual Calibration
F2
Cal. Mode
Y=aX
3. The temperature and the saturation concentration of the calibration cell will be displayed.
Type the volume of mercury which is to be injected and press ENT.
For the first calibration point it is recommended to enter “0” for the volume, in this case
no calibration gas has to be injected.
4. Inject the chosen volume of calibration gas following the steps a-f of the above
description of how using the static calibration source.
5. Press ENT.
After a pause of 10 seconds the GoldTrap is automatically heated and the
measuring signal is recorded.
6. Repeat step 1-3 for additional calibration points.
7. Press F1 if no further calibration points should be measured.
Now the calibration data will be displayed as described in the following.