Operating instructions

4. Calibration
page 30
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
4.2 Show all calibrations
Press F2 in the Calibration Mode menu. A list indicating all previously stored calibrations is
displayed.
The three-letter-code in the right column means:
NOTE: the calibration data table can completely be cleared by pressing F4 !
man
manual calibration
aut
automatic calibration
err
calibration data of an automatic calibration were out of range; in this
case the data of the previous calibration are used.
Allowed range of calibration equation: Y= aX+b; where
0.00002<a<0.00015 and –0.001<b<0.001