Specifications

19
• The LCD displays indicating the tested capacitor is shorted or 
it exceeds the maximum range.
• When capacitance measurement has been completed, disconnect the 
connection between the testing leads and the circuit under test and 
remove the testing leads away from the input terminals of the Meter.
G. Frequency Measurement (See figure 9)
The measurement range is from 10Hz, 100Hz, 1kHz, 10kHz, 100kHz,
1MHz and 10MHz. To measure frequency, connect the Meter as
follows:
1. Insert the red test lead into the
HzΩ terminal and the black 
test lead into the COM terminal.
2. Set the rotary switch to Hz; frequency measurement (Hz) is default 
or press Hz button to select Hz measurement mode.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
• When frequency measurement has been completed, disconnect the 
connection between the testing leads and the circuit under test, and 
remove the testing leads away from the input terminals of the Meter.
• When making frequency measurement at voltage or current range, 
please mind the following signal requirement table:
Range  Signal Requirement  Frequency Range
400mV  ≥ 100mV  1Hz~1kHz
4V  ≥ 0.45V  1Hz~1kHz
40V  ≥ 4.5V  1Hz~1kHz
400V  ≥ 45V  1Hz~1kHz
600V  ≥ 450V  1Hz~200Hz
1000A  ≥ 450A  45Hz~65Hz
Measurement Operation (7)
red
black
Model UT205/UT206: OPERATING MANUAL