Specifications
17
E. Testing Diodes (See figure 7)
Warning
To avoid possible damage to the Meter and to the device under
test, disconnect circuit power and discharge all high-voltage
capacitors before testing diodes.
Use the diode test to check diodes, transistors, and other semiconductor
devices. The diode test sends a current through the semiconductor
junction, and then measures the voltage drop across the junction. A
good silicon junction drops between 0.5V and 0.8V.
To test a diode out of a circuit, connect the Meter as follows:
1. Insert the red test lead into the
HzΩ terminal and the black
test lead into the COM terminal.
2. Model UT205: Set the rotary switch to
, diode measurement
(
) is default or press BLUE button to select measurement
mode.
Model UT206: Set the rotary switch to
Ω and press BLUE
button to select
measurement mode.
3. For forward voltage drop readings on any semiconductor component,
place the red test lead on the component's anode and place the
black test lead on the component's cathode.
The measured value shows on the display.
Note
• In a circuit, a good diode should still produce a forward voltage
drop reading of 0.5V to 0.8V; however, the reverse-voltage drop
reading can vary depending on the resistance of other pathways
between the probe tips.
• Connect the test leads to the proper terminals as said above to avoid
error display. The LCD will display
indicating diode being
tested is open or polarity is reversed. The unit of diode is Volt (V),
displaying the forward voltage drop readings.
Measurement Operation (5)
red
black
Model UT205/UT206: OPERATING MANUAL