Specifications

16
Measurement Operation (4)
• When resistance measurement has been completed, disconnect the 
connection between the testing leads and the circuit under test, and 
remove the testing leads away from the input terminals of the Meter.
D. Testing for Continuity (See figure 6)
Warning
To avoid damages to the Meter or to the devices under test,
disconnect circuit power and discharge all the high-voltage
capacitors before testing for continuity.
To test for continuity, connect the Meter as below:
1. Insert the red test lead into the
HzΩ terminal and the black 
test lead into the COM terminal.
2. Model UT205: Set the rotary switch to
and press BLUE
button to select
measurement mode.
Model UT206: Set the rotary switch to
and press BLUE
button to select
measurement mode.
3. The buzzer sounds if the resistance of a circuit under test is less 
than around 70Ω.
Note
• The LCD displays
indicating the circuit being tested is open.
• When continuity testing has been completed, disconnect the 
connection between the testing leads and the circuit under test, and 
remove the testing leads away from the input terminals of the Meter.
red
black
Model UT205/UT206: OPERATING MANUAL