Specifications

15
Measurement Operation (3)
C. Measuring Resistance (See figure 5)
Warning
To avoid damages to the Meter or to the devices under test,
disconnect circuit power and discharge all the high-voltage
capacitors before measuring resistance.
The resistance ranges are: 400.0Ω , 4.000kΩ , 40.00kΩ , 400.0kΩ ,
4.000MΩ and 40.00MΩ . To measure resistance, connect the Meter
as follows:
1. Insert the red test lead into the
HzΩ terminal and the black 
test lead into the COM terminal.
2. Model UT205: Set the rotary switch to .
Model UT206: Set the rotary switch to
, resistance 
measurement () is default or press BLUE button to select
measurement mode.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
• The test leads can add 0.1Ω to 0.2Ω of error to resistance 
measurement. To obtain precision readings in low-resistance 
measurement, that is the range of 400.0Ω , short-circuit the input 
terminals beforehand, using the relative measurement function button 
REL
to automatically subtract the value measured when the 
testing leads are short-circuited from the reading.
• If Ω reading with shorted test leads is not ≤0.5Ω , check for 
loose test leads, incorrect function selection, or enabled Data Hold 
function.
• For high-resistance measurement (>1MΩ ), it is normal to take 
several seconds to obtain a stable reading.
• The LCD displays
indicating open-circuit for the tested resistor 
or the resistor value is higher than the maximum range of the Meter.