Specifications
15
Measurement Operation (3)
C. Measuring Resistance (See figure 5)
Warning
To avoid damages to the Meter or to the devices under test,
disconnect circuit power and discharge all the high-voltage
capacitors before measuring resistance.
The resistance ranges are: 400.0Ω , 4.000kΩ , 40.00kΩ , 400.0kΩ ,
4.000MΩ and 40.00MΩ . To measure resistance, connect the Meter
as follows:
1. Insert the red test lead into the
HzΩ terminal and the black
test lead into the COM terminal.
2. Model UT205: Set the rotary switch to Ω.
Model UT206: Set the rotary switch to
Ω , resistance
measurement (Ω) is default or press BLUE button to select Ω
measurement mode.
3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
• The test leads can add 0.1Ω to 0.2Ω of error to resistance
measurement. To obtain precision readings in low-resistance
measurement, that is the range of 400.0Ω , short-circuit the input
terminals beforehand, using the relative measurement function button
REL
∆
to automatically subtract the value measured when the
testing leads are short-circuited from the reading.
• If Ω reading with shorted test leads is not ≤0.5Ω , check for
loose test leads, incorrect function selection, or enabled Data Hold
function.
• For high-resistance measurement (>1MΩ ), it is normal to take
several seconds to obtain a stable reading.
• The LCD displays
indicating open-circuit for the tested resistor
or the resistor value is higher than the maximum range of the Meter.
red
black
red
black
Model UT205/UT206: OPERATING MANUAL