Please do not loosen trigger suddenly. As a sensitive
device, Hall element will be sensitive to heat and
mechanical stress to different extents except magnetic
sensitivity. Collision will cause short-term reading
variation.
Please open the clamp head by pressing trigger then
fetch measured conductor by clamp head and loosen
trigger slowly until it is closed completely. Please check if
measured conductor is in the middle of clamp head or not.
Additional error may be caused if failing to place it in the
middle of clamp head. Clamp meter can measure a
current conductor once and measurement reading error
may be caused if measuring two or more current
conductors at the same time.
IX. TECHNICAL INDICATORS
1.General specification
LCD display: 3 3/4-bit LCD display with the maximum
display to 3,999;
Polarity display: Automatic positive/negative polarity
display;
Overload display: To display “OL” or”-OL”; Under-voltage
display:
“ ”
symbol can show that battery voltage is less
than working voltage for battery replacement reference;
Sampling rate: 3 times/second;
Sensor category: Hall affect sensor for DC/AC
measurement;
Error of testing position: ±1.0% of additional reading
error may be caused if failing to place measured source
to center of clamp head during current measurement;
Collision-resistance strength: To bear falling-off collision
for 1m;
The maximum opening dimension of clamp head:
Diameter of 28mm;