●To select functions.
To set DC current measurement to initially set value.
●To connect load.
Please do not loosen trigger suddenly. As a sensitive
device, Hall element will be sensitive to heat and
mechanical stress to different extents except magnetic
sensitivity. Collision will cause short-term reading
variation. Please open the clamp head by pressing
trigger then fetch measured conductor by clamp head
and loosen trigger slowly until it is closed completely.
Please check if measured conductor is in the middle of
clamp head or not. Additional error may be caused if
failing to place it in the middle of clamp head. Clamp
meter can measure a current conductor once and
measurement reading error may be caused if measuring
two or more current conductors at the same time.
10. AC current measurement ( ) (shown in Figure
12)
●To set knob.
To place function knob to “40A “or “600A “measurement
gear.
●To select functions.
To press SELECT key for AC current measurement;
●To connect load.