UT133 Digital Multi meter User Manual I. Introduction UT133A, B are 6000 count palm sized multi meters with auto and manual range respectively. These CE/ETL certified multi meters are CAT Ill 600V, which can withstand 6000kV surge voltage. UT133 are designed with high voltage waning and over range alarm, making this series great for a wide range of measurement needs.
UT133 Digital Multi meter User Manual Safety Instructions er pose a shock hazard 2} Do not use the device if the device or test leads appear damaged o if you suspect ice Ing property. Pay para r attention to the Insulation Briny, ean ee your fingers behind the finger guard. De not input over 800V voltage between the device and the grounding. Use caution to measure voltage >DC 60V or AC victims. Never input voltage and currant exceeding the value listed on the device.
V.General specifications 1) Max voltage between input terminal and earth grounding: 600Vms 2) Fuse Type: 10A Jack: F 10A H 600V Fuse ®6x25mm (or ©6x32mm} map Jack: F 600mA H 600V Fuse $6x32mm 3) Display count: 6000 Overload indication: OL, refresh 3~4 times/s Others: 1) Range: Auto UT133A/Manual UT1338 2) Back light: Manual, shutdown after 30s.
VI. Structure(see figure 1) 1.LCD display 2. Functional buttons 3. Functional dial 4. 10A input terminal 5. COM input terminal 6. Other Input terminals UT133 Digital Multi meter User Manual UT133 Digital Multi meter User Manual VII. Buttons UT133A: * SELECT : Cys switch the functions through AC/DC mV range, frequency, resistance Diode/ continuity, “C'F “REL : The voltage, current and capacitance mode, press this bunion to remove the base. *HOLD/LIGHT : Crass the bunion once to hold the reading.
UT133 Digital Multi meter User Manual VIII. Operation instructions To avoid false reading, replace the battery if the battery low power symbol 55 appears. Also pay special attention fo the warming sign A besides the test lad housing, indicating that the tested voltage or current must not exceed the values listed on the device. 1. AC/DC voltage measurement {see Figure 2) 1) Switch the dial to ACV position 2} Insert the red test lead to COMA jack, black to COM jack.
UT133 Digital Multi meter User Manual A Notes: « If the resistor Is open or over the range, the “OL symbol will ba displayed on the screen. ® Before measuring resistance, switch off the power supply of the circuit, and fully discharge all capacitors. # When measuring low resistance, the test leads will produce measurement armor. To obtain accurate measurement, short the test lads and use REL function. # If the resistance when shorted is more than 0.50, please check if test leads are loosened or damaged.
UT133 Digital Multi meter User Manual 5. Capacitance measurement (see figure 3) 1) Switch the dial to capacitance measurement insertions the red test lead to Velma jack, black to COM jack. 3)Reading Is displayed. A Notes: ® Switch off the power supply to the circuit, and fully discharge all capacitors # Before measuring capacitors (especially for high voltage capacitors), please fully discharge them.
UT133 Digital Multi meter User Manual 8. AC/DC current measurement (AC current measurement Is only for UT133A) 1)Switch the dial to AC/DC current position 2)According to the current being measured. Insert the red test lead to Velma jack or 10A jack, black to COM jack. 3)Connect test leads with the circuit In series 4)Reading is displayed. A Notes: ‘# Ba fore measuring, switch off the power supply of the cultural.
7. Battery measurement (Only for UT133B) 1) Switch the dial to battery position 2) Insert the red test lead to VOA Jack or 10A jack, black to COM jack. 3) Connect red test lead to + pole of the battery, black to — pole. 4) Reading is displayed ( Good-normal status; Low-low power; Bad-ends replacement) 5) LCD display of battery status UT133 Digital Multi meter User Manual Figure 5 UT133 Digital Multi meter User Manual 1.5V battery: resistance=300 Indication Voltage Good =131 Low So Bad <0.
8. Temperature measurement (Only for UT133A) 1) Switch the dial to temperature position 2)Insert K-type thermonuclear to the device and place the test probes on the object under measurement. 3)Reading Is displayed. A Note * Only K-types thermonuclear is applicable. *® The measured temperature should be less than © Tum on the device, after "OL" symbol appears, insert K-thermonuclear into the device.
UT133 Digital Multi meter User Manual UT133 Digital Multi meter User Manual 8. NCV measurement (ses figure 7) 1) Switch the dial to NCV position 2) Place the device near the measured object. indicates the intensity of the galactic fled. More "and the higher the buzzer frequency, the higher the electric field intensity. 3) Intensity of electric field.
UT133 Digital Multi meter User Manual IX. Others ® The device enters measurement status in 2 seconds after startup. Restart the device ripened" appears. ® The device automatically shuts down If there Is no operation for 15 minutes. You can wake up the device by pressing any key. To disable auto shutdown: switch the deal io OFF position and long press DELETE button until the device tum on.
9. AC current (Only for UT133A) UT133 Digital Multi meter User Manual Range Position Model Resolution Accuracy 600.0pA 0.14A Fr din 02 600.0mA outta 0.1mA BA 0.001A 1.5% + 10.00A 0.01A 1.5%) Frequency response: 40~400Hz Display: true RMS Accuracy guarantee range: 5-100% of the range, shorted circuit allows least significant digit<2 Input buzzer goes off; input “OL” symbol appears Overload protection: {similar to DC current) 3 UT133 Digital Multi meter User Manual XI.
UT133 Digital Multi meter User Manual Fuse replacement: 1) Switch the dial to “OFF “position and remove the test leads from the Input terminal 2) Loosen the both screws on the rear cover, and then remove the rear cover to replace the fuse Fuse specification F1 Fuse $6x32mm F 600mA H 600V F2 Fuse ®6x25mm (or ®8x32mm) F 10A H 600V Figure 8b UT133 Digital Multi meter User Manual Test probe replacement If Insulation on prob Is damaged, replace It.