Manual

106
DXi User Manual
Appendix B: Flow Sensors
Flow Sensor Installation Instructions
The plastic cover of the DXi main board shows all inputs and terminal polarity.
To install a Irritrol or Toro flow meter:
1. Power down the DXi controller.
2. Connect the white and black wires as shown.
3. Return power to the controller.
4. To configure the flow sensor, go to Setup-->Flow-->Flow Sensors.
See Chapter 4: Setup on how to navigate the Flow Sensors screen.
Use the "Flow Meter Offset and K Values" tables (next page) to properly
calibrate the flow sensor.
For complete installation and setup instructions of a Flow decoder, refer to Toro
document 373-1022, TW-DAC-FLOW Decoder Installation Guide.
Flow Sensor Overview
To establish flow limit checking either on a controller or individual station basis,
verify the following:
Flow meter is installed properly.
Correct Offset and K values are entered.
Station Upper Limits have been established. See "Max Flow Limit".
Station Lower Limits have been established. See "Min Flow Limit".
Main line flow limits have been established. See "Main Flow".
Total Monthly Flow has been set. See "Flow Options".
Unscheduled Flow Limit has been defined. See "Unscheduled Flow Limit".
Upper and lower limit checking is enabled.
See "Enable/Disable Limit Checking".
The appropriate flow limit check delay has been established. See "Delay Limit".
Selection of at least one of the three flow meters.
FLOW 1”
MV1 MV2 MV3 P1 P2
24
VAC COM COM COM
FLOW/SENSORS/ALARM
1 2 3 1 2 3 4 5 6
1 2 3 1 2
OFF
FLOW AUX.SENSOR INPUTS MV PUMP
+ - + - + - + - + - + - + - + - + - + + + + + - - -
2 WIRE
24
VAC
F2
5V
12V
3.5V
24VAC
COMMON
DX3 MAIN BOARD
white to +
black to -
CST flow meter