User manual
62
Specications
4
Specications
Sources all analog channels
Mask definition Mask enclosing acquired waveform
with user-defined tolerance
Actions
on mask violations beep, acquisition stop, screenshot,
trigger pulse, automatically saving trace
data
during acquisiton Statistics: number of completed tests
(max. 4x10
-9
events), number of
passes / failed acquisitions (absolute
and in percent), test duration
Waveform maths
Quickmath
Functions addition, substraction, multiplication,
division
Sources 2 analog channels
Mathematics
Functions addition, substraction, multiplication,
division, minimum / maximum, square,
square root, absolute value, pos/neg
wave, reciprocal, inverse, log10/ln,
derivation, integration, filter (lowpass/
highpass)
Editing formula editor, menu-driven
Sources all analog channels, user-defined
constants
Storage location math. memory
Number of formula sets 5 formula sets
Number of equations 5 equations per formula set
Simultaneous display of math.
functions
1 formula set with max. 4 equations
Frequency Analysis (FFT)
Parameters frequency span, center frequency,
vertical scale, vertical position
FFT length 2 Kpts, 4 Kpts, 8 Kpts, 16 Kpts, 32 Kpts,
64 Kpts
Window Hanning, Hamming, Rectangular,
Blackman
Scale dBm, dBV, V
rms
Waveform arithmetics refresh, envelope, average (up to 512)
Cursor measurement 2 horizontal cursors, previous/next peak
search
Sources all analog channels
Pattern Generator
Functions square wave / probe adjust, bus signal
source, counter, pseudo-random
pattern
Square wave
(Probe ADJ output)
frequencies: 1 kHz, 1 MHz
level: 1 V
pp
(t
a
<4 ns)
Bus Signal Source (4 Bit) I
2
C (100 kBit/s, 400 kBit/s, 1 MBit/s), SPI
(100 kBit/s, 250 kBit/s, 1 MBit/s), UART
(9600 Bit/s, 115,2 kBit/s, 1 MBit/s)
Counter (4 Bit) frequencies: 1 kHz, 1 MHz
direction: decrementing
Random pattern (4 Bit) frequencies: 1 kHz, 1 MHz
Component Tester
Parameters voltage (X), current (Y)
Testing frequency 50 Hz, 200 Hz
Voltage 10 V
p
(open)
Current 10 mA (short)
Reference potential Ground (PE)
Interfaces
Connectors and ports
for mass storage (FAT16/32) 2 x USB-Host (Typ A), max. 500 mA
for remote control HO720 dual interface:
USB Device (Typ B), RS-232
Optional interfaces HO730 dual interface:
USB Device (Typ B), Ethernet (RJ45)
HO740 interface: IEEE-488 (GPIB)
External monitor interface DVI-D (480p, 60Hz) HDMI compatible
Y-OUT (BNC) Trigger event on mask violation:
pulse:>0.5μs
General Data
Application memory 4 MB for references, formulas, device
settings, language and help functions
Save/Recall
device settings
on internal file system or external USB
memory, available file format: HDS
reference waveforms on internal file system or external USB
memory, available file formats: BIN,
CSV, TXT, HRT
traces on external USB memory, available file
formats: BIN, CSV, TXT, HRT
data display or acquisition data
sources single or all analog channels
screenshots on external USB memory, available file
formats: BMP, GIF, PNG
(color, inverted, grey-scale)
Math equation sets on internal file system or external USB
memory
Realtime Clock (RTC) date and time
Power supply
AC supply 100 V to 240 V, 50 Hz to 60 Hz, CAT-II
Power consumption
2-channel models max. 45W
4-channel models max. 55W
Safety in line with IEC 61010-1 (ed. 3), IEC
61010-2-30 (ed. 1), EN 61010-1, EN
61010-2-030 , CAN/CSA-C22.2 No.
61010-1-12 , CAN/CSA-C22.2 No.
61010-2-030-12 ,UL Std. No. 61010-1
(3rd Edition) , UL61010-2-030
Temperature
Operating temperature range +5 °C to +40 °C
Storage temperature range -20 °C to +70 °C
Rel. humidity 5% to 80% (without condensation)
Mechanical Data
Dimensions (W x H x D) 285 x 175 x 140 mm
Net weight 2.5 kg
All specifications at 23°C after 30 minutes warm-up.
5
Specications
Accessories included:
HO720 RS-232/USB dual interface, line cord, printed opera-
ting manual, software-CD, 2/4 probes (model depending):
HZ154 (up to 100 MHz, 10:1/1:1 switchable) or HZO10 (up to
250 MHz, 10:1)
Accessories:
HO3508 8 channel logic probe (350 MHz, 4 pF)
HZ115 Differential Probe 100 : 1/1000 : 1
HZO20 High voltage probe 1000 : 1 (400 MHz,
1000 V
rms
)
HZO30 1GHzactiveprobe(0.9pF,1MΩ)
HZO40 Active differential probe 200 MHz (10 : 1, 3.5 pF,
1 M Ω )
HZO41 Active differential probe 800 MHz (10 : 1, 1 pF,
2 0 0 k Ω )
HZO50 AC/DC Current Probe 30 A, DC to 100 kHz
HZO51 AC/DC Current Probe 100/1000 A, DC to 20 kHz
HZ51 150MHzpassiveprobe10:1(12pF,10MΩ)
HZ52 250MHzpassiveprobe10:1(10pF,10MΩ)
HZ53 100MHzpassiveprobe100:1(4.5pF,100MΩ)
HZO90 Carrying case for protection and transport
HZO91 4 RU 19” rackmount kit