User manual

31
Measurements
PEAK VALUES: This mode provides two cur-
sors to measure the minimum and the maxi-
mum voltage of a signal within the interval set
by using both cursors. The values Vp- and Vp+
correspond to the minimum and the maximum voltage.
The peak value (Vpp) corresponds to the amount of vol-
tage between the minimum and maximum value.
RMS, MEAN, STANDARD DEVIATION: This
mode provides two cursors to measure the ef-
fective value (RMS – Root Mean Square), the
mean value and the standard deviation within the interval
set by using both cursors.
DUT Y RATIO: This mode provides three cur-
sors to determine the duty ratio between the
two horizontal cursors. The third cursor is
used to specify the threshold at which the duty ratio is
measured.
RISE TIME 90%: This mode provides two cur-
sorstomeasuretheriseandfalltimeoftherst
rising slope in the displayed view of the screen
within the time span set by the cursors. The rise time iden-
tiesthetimeinwhichthesignalrisesfrom10%to90%of
its amplitude.
RISE TIME 80%: This mode provides two cur-
sorstomeasuretheriseandfalltimeoftherst
rising slope in the displayed view of the screen
within the time span set by the cursors. The rise time iden-
tiesthetimeinwhichthesignalrisesfrom20%to80%
of its amplitude.
V-MARKER: This mode pro-
vides two cursors to measure
three different voltages and a
time. The values V1 and V2 correspond to the voltage bet-
ween the zero base line of the selected curve and the cur-
rentpositionoftherstorsecondcursor.ThevalueΔV
corresponds to the amount of voltage between the two
cursors.ThevalueΔtcorrespondstotheamountoftime
between the two cursors.
To move cursors, press the universal knob in the CURSOR/
MENU section and position the cursor by turning the univer-
sal knob. The measurement types have the following func-
tions. The measurement results are displayed at the bottom
of the screen. If “n/a” is displayed, the measurement is not
applicable to the signal. For instance, this may be the case
for a voltage measurement on a POD because only logic sta-
tes without voltage reference are displayed here. If “?” is dis-
played, the display does not show a complete measurement
result. For instance, the period to be measured may not dis-
playcompletelyandcanconsequentlynotbeidentied.
If the function AUTOM. SOURCE is activated (ON), the
currently targeted channel will be used as source for the
measurement automatically. If the setting is deactiva-
ted (OFF), the channel set under SOURCE will be applied
even if it is not targeted. Pressing the soft menu key SET
TO TRACE places the selected cursors in their optimal po-
sition on the signal curve. This allows very fast and typi-
cally optimal automatic positioning of the cursors. For the
mostpart,onlynetuningisrequiredatthispointandthe
tedious major adjustments to the cursors will no longer be
necessary. In case the automated function SET TO TRACE
does not provide the anticipated results due to complex
signals, you can press the key SET TO TRACE to posi-
tionthecursorsinapredenedstartingposition.Thisal-
lows you to return distant cursors to the screen. The soft
menu key GLUE TO TRACE allows cursors to stay on the
selected data point without changing the position in the
measurementsignalevenifthescalingismodied(cur-
sors will be „glued“ to the signal). This function can be ac-
tivated or deactivated. If this mode is deactivated, the cur-
sor stays in position on the screen if scaling occurs. With
GLUE TO TRACE deactivated, the measured value changes
whileitremainsunmodiedwhenthemodeisactivated.
The cursors will be deactivated by pressing the CURSOR
MEASURE key. The cursor menu will be closed.
8.2 Automatic Measurements
The R&S®HMO1002 features cursor measurements and
also various automatic measurements. These may be acti-
vated by pressing the key AUTO MEASURE
11
in the sec-
tion ANALYZE. This menu allows you to select up to six
automatic measurement functions by using the soft menu
key MEAS. PLACE and the universal knob in the CURSOR/
MENU section. Following automatic measurement types
can be selected:
MEAN VALUE: This mode measures the mean
value of the signal amplitude (Mean). If the sig-
nalisperiodic,therstperiod on the left of the
screen will be used for the measurement. The measure-
ment will only be applied to the selected channel.
Fig. 8.1: Cursor measurements selection menu