Datasheet
Technical data(Pt100Ω) for FS-400P
RESISTANCE VS TEMPERATURE RELATIONSHIP OVER THE RANGE -200 TO + 600
FOR PLATINUM RESISTANCE TEMPERATURE DETECTOR ELEMENTS
emf/µV M : DIN
(t90) 0123456789(t90)
-200
18.52
-200
-180 27.10 26.67 26.24 25.82 25.39 24.97 24.54 24.11 23.68 23.25 -180
-160 35.54 35.12 34.70 34.28 33.86 33.44 33.02 32.60 32.18 31.76 -160
-140 44.38 43.36 43.05 42.63 42.22 41.80 41.39 40.97 40.56 40.14 -140
-120 52.11 51.70 51.29 50.88 50.47 50.06 49.65 49.24 48.83 48.42 -120
-110 56.19 55.79 55.38 54.97 54.56 54.15 53.75 53.34 52.93 52.52 -110
-100 60.26 59.85 59.44 59/04 58.63 58.23 57.82 57.41 57.01 56.60 -100
-90 64.30 63.90 63.49 63.09 62.68 62.28 61.88 61.47 61.07 60.66 -90
-80 68.33 67.92 67.52 67.12 66.72 66.31 65.91 65.51 65.11 64.70. -80
-70 72.33 71.93 71.53 71.13 70.73 70.33 69.93 69.53 69.13 68.73 -70
-60 76.33 75.93 75.53 75.13 74.73 74.33 73.93 73.53 73.13 72.73 -60
-50 80.31 79.91 79.51 79.11 78.72 78.32 77.92 77.52 77.12 76.73 -50
-40 84.27 83.87 83.48 83.08 82.69 82.29 81.89 81.50 81.10 80.70 -40
-30 88.22 87.83 87.43 87.04 86.64 86.25 85.85 85.46 85.06 86.46 -30
-20 92.16 91.77 91.37 90.88 90.59 90.19 89.80 89.40 89.01 88.62 -20
-10 96.09 95.69 95.30 94.91 95.52 94.12 93.73 93.34 92.95 92.55 -10
0 100.00 99.61 99.22 98.83 98.44 98.04 97.65 97.26 96.87 96.48 0
0 100.00 100.39 100.78 101.17 101.56 101.95 102.34 102.73 103.12 103.51 0
10 103.90 104.29 104.68 105.07 105.46 105.85 106.24 106.63 107.02 107.40 10
20 107.79 108.18 108.57 108.96 109.35 109.73 110.12 110.51 110.90 111.29 20
30 111.67 112.06 112.45 112.83 113.22 113.61 114.00 114.38 114.77 111.15 30
40 115.54 115.93 116.31 116.70 117.08 117.47 117.86 118.24 118.63 119.01 40
50 119.40 119.78 120.17 120.55 120.94 121.32 121.71 122.09 122.47 122.86 50
60 123.24 123.63 124.01 124.39 124.78 125.16 125.54 125.93 126.31 126.69 60
70 127.08 127.63 127.84 128.22 128.61 128.99 129.37 129.75 130.13 130.52 70
80 130.90 131.28 131.66 132.04 132.42 132.80 133.18 133.57 133.95 134.33 80
90 134.71 135.09 135.47 135.85 136.23 136.61 136.99 137.37 137.75 138.13 90
100 138.51 138.88 139.26 139.64 140.02 140.40 140.78 141.16 141.54 141.91 100
110 142.29 142.67 143.05 143.43 143.80 144.18 144.56 144.94 145.31 145.69 110
120 146.07 146.44 146.82 147.20 147.57 147.95 148.33 148.70 149.08 149.46 120
130 149.83 150.21 150.58 150.96 151.33 151.71 152.08 152.46 152.83 153.21 130
150 157.33 157.70 158.07 158.45 158.82 159.19 159.56 159.94 160.31 160.68 150
180 168.48 168.85 169.22 169.59 169.96 170.33 170.70 171.07 171.43 171.80 180
200 175.86 176.22 176.59 176.96 177.33 177.69 178.06 178.43 178.79 179.16 200
300 212.05 212.41 212.76 213.12 213.48 213.83 214.19 214.54 214.90 215.25 300
400 247.09 247.44 247.78 248.13 248.47 248.81 249.16 249.50 249.85 250.19 400
600 313.71 314.03 314.35 314.67 314.99 315.31 315.64 315.96 316.28 316.60 600
* Applicable FOX- Model Name : FS-400P / FS-401P / FS-402P
Originally its element can measures from -200°C to more than 600°C, but we would like to
recommend the operating range with stability as follows ;
* recommendable operating range : -200 ~ + 200°C
* Therefore our 3 models can measures above operating range with stability.
However, please use these models proper method without any damage or injury of its tube,
& wire when it is used at high temperature
.
And with regard to theirs size and shape, please refer to our products catalogue.