Installation guide
91
Table 2 (at page 18). HVT is based on UMX APIs. Please make sure the RCB driver and the
HwTest driver are properly installed.
11.4.1 The Sine Wave Test
a) Start the test.
Set up two Sora boxes. Place the two antennas about one meter apart.
Select sine test on both boxes.
Choose one box as the sender (select send) and the other as the receiver (select
receive).
Click start button (b) at the sender. You will see the sender’s status bar (j) showing
“Sending 1MHz sine wave”. If a failure occurs, the error message will be shown in the
log window (k).
Select a proper Tx gain (t) setting at the sender (for example 15dB).
Choose the same sampling rate (r) as your RAB at the receiver.
Click start button (b) at the receiver to capture the signal. You will see the waveform
displayed on two popup windows, as shown in the first row of Table 7. If failures occur,
the error messages will be shown in the log window (k).
By default, the receiver gain is automatically adjusted (The AGC box (p) is checked). But
the user can disable the AGC and adjust the receiver gain manually by unchecking the
box.
The user can also start a central frequency offset (CFO) calibration by clicking the auto
calibration button (e). The tool will automatically compute the frequency offset
between the sender and receiver and adjust the receiver’s frequency setting to match
the sender’s. The calibration result will be shown in the log window (k).
The calibration results (parameters) can be stored in a file and the user can late
configure the RF front-end based on these values. Figure 41 shows a sample
configuration file.
[Parameter]
freqOffset=8064
txgain=0xb80
rxgain=0x1800
rxpa=0x1000
centralFreq=2422
sampleRate=40










