Instruction Manual

Forehead Sleep Recorder File No. DMR-SM902-4-25 Ver:A0
11 / 23 Copyright© Shenzhen BrainScience Bio-Info Technology Co., LTD
Before each use of this product, the smart mobile device or tablet running the APP for the
product should be fully charged to ensure that there is enough power to support the operation of
the APP during the recording. At the same time, the APP should be set to the locked state to
prevent accidental exit and avoid incomplete recording data or data corruption.
Step 1: Before wearing the device, please clean subject' skin and wash away cosmetics, dirt and
oil stains to prevent the device's adhesion from falling, and signal disruption.
Step 2: Take out an electrode sticker and install it onto the device tightly.
Warning: Please purchase the electrode specified by the company, otherwise the accuracy of
the signal and safety cannot be guaranteed, or it may cause allergic, irritating and other adverse
reactions.
Step 3: Tear off the release film on the back of the electrode sticker to expose the gel part.
Step 4: Press the power button on top of the device, and the green light is always on. Here the
device is in the startup state.
Step 5: With button on top side, align the device with the center of the forehead and attach it
tightly, (be careful not to stick it to the hair).
The product should be fully charged immediately after each use, In order to facilitate the use
of this product next time.
7.4.1. Descriptions for indicator lamp
Table 6 -Descriptions for device indicator lamp status
Power status
Battery status
Bluetooth status
Indicator lamp status
Charging
<100%
Any status
Yellow light breathing (5s
stop, 5 breathing)
100%
Steady green light
Stand by and
Not charge
>5%
Connected
Any status
Not connected
Steady green light
0~5%
Any status
Yellow light flashes
quickly
Shut down
without charging
Any status
off
Step1 Step2 Step3 Step4 Step5