Unit installation
22
Ballast connectors
more
ballast replacement
Ballast replacement is not part of the UV unit’s routine
maintenance. However, in the event that a ballast needs to
be replaced, the following procedure should be followed:
1. Power down the UV unit by turning the unit isolator
switch, if installed, to the “OFF” position. If not, ensure
that power to the unit is removed by opening the switch
or breaker upstream of the UV unit.
2. Locate the old ballast to be replaced. Refer to the
electrical diagram, if necessary.
3. Isolate the old ballast from rest of system by disconnecting
the ballast connector.
4. Using 5/32 L-shaped hex wrench completely remove
the bolt securing the ballast to the sub panel. Put the
bolt aside.
5. Remove the old ballast by pulling it towards you.
Discard the old or defective ballast.
6. Install the new ballast, securing it with the bolt
removed earlier.
7. Tighten the bolt with 5/32 L-shaped hex wrench.
8. Reconnect ballast connector(s).
9. Restore the power to the UV unit by turning the unit
isolator switch to the “ON” position or by closing the
upstream switch or breaker.