Specifications

Table Of Contents
Version 1.1 rev Jan 2013
Page 94
iXon Ultra
Hardware
4.7 - SIGNAL DIAGRAMS
iXon Ultra Input & Output Timing Hardware
Figure 42: External Trigger
External Trigger Input (at connector)
• VIH(Highlevelinputvoltage,minimum)=2.2V
• VIL(Lowlevelinputvoltage,maximum)=0.88V
• Changetofallingedgetriggerbyconnecting“TriggerInvertInput”pintoGroundwith≤500Ω
Figure 43: Fire & Shutter
Fire, Shutter, Arm & Frame Output (at connector)
• Use50Ωcable+highimpedanceinput(>1kΩ)
• DrivebothTTL&CMOSunterminatedinput
4.8 - CAMERALINK
The iXon Ultra is equipped with a Base Conguration (3-tap interface) Camera Link output which conforms to the
specication dened by the Automated Imaging Association (AIA). This provides access to the camera data output with
very low latency. Note that this is an OUTPUT ONLY.
A Technical Note “Camera Link Output” is available at:
http://www.andor.com/learning-academy/camera-link-output-ixon-ultra-output-for-direct-data-access