Instruction manual

Operation
Print utilities (option)
Thermo Scientific WX Ultra Series 2-81
(4) Zonal speed
This function can change (in units of 100 rpm) the zonal speed in the range of 2,000 to 3,000 rpm.
For the zonal speed, see Section Zonal operation)
Figure 2-29
Run condition setting screen
4. Select option unit
This function allows you to set one of the following optional units through the RS232C.
1) Printer
2) Compass Centrifugation Software
3) Unused
Figure 2-30
Optional Unit connection screen
Print utilities (option)
This submenu option is used to print the various information maintained in the ultracentrifuge on the
optional printer. The information includes run records, rotor usages, memory contents, etc. When you
select “5. PRINT UTILITIES” from the “user customization routines” submenu, this submenu option
starts working and displays the PRINT UTILITIES screen.
Note When “Unused” is selected in the optional unit setting, machine log data will be deleted in
the order starting with the oldest data and new data will be stored in the memory if the machine log
data exceeds the memory area capacity of the centrifuge. For managing such machine log data, be
sure to set an optional unit. When such an optional unit has been set, the printer printout or PC
data input request message is displayed if the machine log data exceeds the memory area capacity.