Instruction manual
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-57
Step Key operation Screen display and considerations
1 Set the Speed in the Run Screen. >> See Section Setting run conditions.
2 (1) Use cursor keys to move the cursor
to RCF • w
2
T on the Run Screen and
press the ENTER key.
(2) Use cursor keys to move the cursor
to w
2
T and press the ENTER key.
3 Use cursor keys to move the cursor to
the w
2
T setting field and enter a
desired w
2
T value.
Example: 9.87 x 1011
4 When you press START key, the
machine starts to perform an w
2
T run.
• The contents of the display
in the FUNCTION field
change.
• ω
2
T blinks.
• The FUNCTION field
switches to the ω
2
T field.
ω
2
T display field
ω
2
T setting field
• The value you entered is
displayed in the ω
2
T setting
field.
• The time value is calculated
from the speed and the ω
2
T
value and the run time is
displayed in the Time Setting
field.
• Arithmetic and integral
operations are performed in
the ω
2
T display field.
ω
2
T display field