Instruction manual
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-53
How to set an RCF value
Enter a rotor number and an RCF value, and the machine will calculate, set, and display the speed.
Shown below is the procedure.
Step Key operation Screen display and considerations
1 While in the Run Screen, set a speed. >> See Section Setting run conditions.
2 (1) While in the Run Screen, use cursor
keys to move the cursor to RCF • ω
2
T
and press ENTER key.
(2) Move the cursor to RCF
max
(or
RCF
avg
) and press ENTER key.
3 Check that ROTOR LIST blinks and
press the ENTER key.
• The FUNCTION field
switches to what is shown
in the diagram to the left.
•The RCF
max
field blinks.
•The FUNCTION field
switches to the RCF field.
• The ROTOR LIST field
blinks.
• The display switches to the
Choose Rotor screen.