Instruction manual

Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-53
How to set an RCF value
Enter a rotor number and an RCF value, and the machine will calculate, set, and display the speed.
Shown below is the procedure.
Step Key operation Screen display and considerations
1 While in the Run Screen, set a speed. >> See Section Setting run conditions.
2 (1) While in the Run Screen, use cursor
keys to move the cursor to RCF • ω
2
T
and press ENTER key.
(2) Move the cursor to RCF
max
(or
RCF
avg
) and press ENTER key.
3 Check that ROTOR LIST blinks and
press the ENTER key.
The FUNCTION field
switches to what is shown
in the diagram to the left.
•The RCF
max
field blinks.
•The FUNCTION field
switches to the RCF field.
The ROTOR LIST field
blinks.
The display switches to the
Choose Rotor screen.