Instruction manual
Operation
How to use the FUNCTION field
Thermo Scientific WX Ultra Series 2-39
(2) Deleting a program
This section explains how to delete a set of run conditions stored in program memory.
To delete a memory number, delete all steps in that memory number.
Note While the instrument is running, you cannot delete a program memory number.
Step Key operation Screen display and considerations
1 While in the Run Screen, use cursor
keys to move the cursor to PROG • RTC
and press ENTER key.
Then move the cursor to PROG and
press ENTER key.
2
Press the upward cursor key to
cause PROGRAM to blink, then press
ENTER key.
3 Use numeric keys to enter the memory
unit number you want to delete.
To delete memory unit 3, press
4 Select “Deleting”.
• The FUNCTION field switches
to the PROGRAM field.
• The MEMORY No. setting field
blinks.
• The display switches to the
Program Map screen and the
MEMORY No. field blinks.
• The image of the memory unit
you have just specified appears.
• The message field in the bottom
row switches to what you see on
your left.