Instruction manual

Operation
How to use the FUNCTION field
2-26 WX Ultra Series Thermo Scientific
6 To select a desired rotor from the
“Rotor Management”:
(1') Select “Rotor Management”.
(2') Select a desired rotor using the
upward or downward cursor key
.
(When using fixed angle rotor T-8100)
(3') Select “Yes”.
To select no rotor:
Select “None”.
7 Enter values into SPEED (RCF
max
or
RCF
avg
or ω
2
T), TIME, TEMP, A
(ACCEL), and D (DECEL).
>> To enter run conditions in multiple steps in a step-mode operation, continue
the operation from the above. (See Section Step-mode operation.)
>> You cannot skip a step. (You cannot enter anything in step 2 with nothing
entered into step 1.)
The display switches to the Rotor
Management screen and the
Rotor List is displayed.
The rotor type you have just
entered appears in the top row in
the screen.
The SPEED field in the SPEED
priority screen blinks.
The SPEED field in the SPEED
priority screen blinks.
Example
(Downward cursor key)
When you finish entering run
conditions for step 1, step 2 enters
into an input wait state.