Instruction manual
Operation
Basic operation
2-10 WX Ultra Series Thermo Scientific
Using Rotor Life Management
The rotor type and serial number may be set in rotor management to allow for rotor logging within the 
centrifuge. Whether the rotor type and serial number for a particular rotor have been set or not, it does 
not affect the normal operation of the ultracentrifuge. However, you are advised to set the rotor type 
and serial number of the rotor for each run to allow effective rotor life management. Any registered 
rotor that does not have its rotor type and serial number set will appear as an undefined rotor in the 
displayed list (“Undefined rotor 1”). 
You can use a particular rotor as undefined if you need to manage the rotor separately from the other 
defined rotors in terms of the number of runs made and accumulated run time (“Undefined rotor 2”). 
For more information on rotor life management, see Section Rotor management.
Step Key operation Screen display and considerations
1 On the Run Screen, use cursor keys to 
move the cursor to RLM and press the 
ENTER key.)
2 On the rotor control field, use the left 
and right arrow keys
 and   to select each page.
3 Use the up and down arrow keys to 
move the cursor to the row of the rotor 
type to be used and press the ENTER 
key.
In this example, the cursor moves to 
“T-8100-1234”.
Select “Running”.