Thermo Scientific WX Ultra Series 2-5
[run screen]
The screen for displaying run conditions and operational status is called the run screen.
Speed, time, and temperature are displayed in two rows: the top row displays the current actual run 
conditions, while the bottom row displays the set run conditions.
The acceleration (ACCEL) and the deceleration (DECEL) fields display set conditions.
Figure 2-3 
Run screen
[Cursor key]
Pressing the cursor key will highlight the field where changes can be made. (This blinking/highlighted 
object is referred to as the cursor in this manual.)
The screen setting field is in either of the following states depending on whether the cursor is there.
1. Determined input state: This is a normal state and the cursor does not appear.
2. Input wait state: Press a cursor key (either the top, bottom, right, or left) while in the determined 
input state, and the numerical part of the setting field will blink a 0 (or numerical value) and 
display the cursor. In this state, the system accepts a numerical input. Press cursor keys to move the 
cursor.
To set a run condition, enter the cursor into an input wait state, move the cursor to the item you want 
to set, then enter a value. If you have made no keystroke (such as a numerical input) for more than 30 
seconds, the system will automatically enter a determined input state.
Note When the screen is in a determined input state without a cursor (when the Run Screen is on 
after power-up), if you wish to enter a numerical value in a specified parameter field, press a cursor 
key (either of the upward, downward, right, and left arrow keys), move the cursor to the specified 
parameter field, then enter the value.
The cursor keys display and move a cursor. Once a cursor appears, pressing a cursor key moves the 
cursor to the corresponding direction (upward, downward, rightward, or leftward).
Operation state 
display field
Setting value 
display field
Run mode indicator
}
}
}