Instruction manual
Description
Design
Thermo Scientific WX Ultra Series 1-7
Safety devices
(1) Containment
Should a rotor failure occur, the guard ring will contain the contents of the rotor inside the centrifuge 
(Fig. 1-4).
(2) Imbalance detector
If during operation the vibration of the rotor becomes excessive due to serious imbalance or improper 
bucket setting, the imbalance detector detects the situation and decelerates the rotor immediately. 
However, the ultracentrifuge is designed to tolerate imbalance associated with visual balancing-it is 
equipped with an imbalance tolerant drive. (For more information on the balancing of rotors, see 
Section 2-1-2, “Preparing tubes/bottles and rotor”.)
(3) Door lock system
The chamber door automatically locks for safety while the rotor is spinning. When the power supply is 
off, the door remains locked. The door can only be opened and closed when the rotor is at rest and the 
rotor chamber is vented. Unless the door is closed, the rotor will not start rotating except in zonal 
mode. To open the door in the event of a power failure, see Section 2-7, “When power failure occurs”.
(4) Speed sensor and overspeed detector
For protection in the event of entry errors the ultracentrifuge is provided with an automatic system to 
stop the rotor when its speed exceeds the maximum allowable speed. If a speed higher than the 
maximum permitted speed is set, the ultracentrifuge will detect the mistake before the speed reaches 
3000 rpm, and then will display an alert message and decelerate the rotor to a stop.