Instruction manual

Thermo Scientific WX Ultra Series 6-1
6
Specifications
The Sorvall WX Ultra series centrifuges are manufactured and tested according to the following
regulations of EMC:
EN 61326
EN 61000-3-2
Model Sorvall WX Ultra Series
Sorvall WX 100 Ultra Sorvall WX 90 Ultra Sorvall WX 80 Ultra
Maximum speed 100,000rpm 90,000rpm 80,000rpm
Maximum RCF 802,006 x g
(T-8100)
692,149 x g
(T-890)
602,644 x g
(T-880)
Speed control accuracy ±10 rpm (1,000 rpm to maximum speed)
Acceleration/deceleration
control
About 5 minutes from 0 to 100,000 rpm or vice versa (P100AT2)
Rotor temperature
control/display accuracy
±0.5°C set temperature is from 0°C to 40°C
Set speed 1,000 rpm to maximum speed in increments of 100 rpm
Vacuum system Oil rotary vacuum pump and oil diffusion pump combined
Ultimate vacuum: below 0.01Torr (1.33 Pa) (This value reaches within 20
minutes after the rotor starts rotating.)
Noise level 53 dB (A scale) (measured 1m in front of the instrument at the set speed
of the rotor)
Maximum heat
dissipation into room
1kW/hr
Cooling method Thermo-module cooling (CFC-free)
User interaction means Color liqiud crystal display (color-256), touch panel, and keyboard
Interface RS232C X 1CH
Dimensions Width:790 mm; depth:690 mm; height:1,000 mm
Depth with safety cover mounted:890 mm
Height to table:850 mm
Weight 400 kg
Power requirement Single phase 50/60 Hz
208,220 Vac+/-10 %, 20A maximum (normally 8A)
230,240 Vac+/-10 %, 16A maximum (normally 7A)
Environment conditions Ambient temperature for operation: 2°C to 40°C
With maximum relative humidity of 80% at 31°C, linearly decreasing to 50% at 40°C.
Ambient Temperature for performance guarantee: 10°C to 30°C
The unit is intended for indoor use only
Maximum altitude of 2000 meters.
Pollution degree 2
Overvoltage category II