User`s guide
Appendices
APPENDIX A: SYRINGE VOLUME/DIAMETER REFERENCE TABLE
H
arvard Apparatus
S
tainless Steel
Size Diameter
2.5 ml 4.851 mm
8 9.525
20 19.13
50 28.6
100 34.9
Air–Tite
HSW Norm-Ject
Size Diameter
1. ml 4.69 mm
2.5 9.65
5 12.45
10 15.9
20 20.05
30 22.9
50 29.2
Becton Dickinson
‘Plasti-pak’
Size Diameter
1. ml 4.699 mm
3 8.585
5 11.989
10 14.427
20 19.05
30 21.59
50 26.594
60 26.594
Terumo
Size Diameter
1. ml 4.70 mm
3 8.95 mm
5 13
10 15.8
20 20.15
30 23.1
60 29.1
S
GE Scientific Glass
E
ngineering
Size Diameter
5. µl 0.343 mm
10 0.485
25 0.728
50 1.03
100 1.457
250 2.303
500 3.257
1. ml 4.606 mm
2.5 7.284
5 10.301
10 14.567
25 23
50 27.5
100 35
Hamilton
Glass - All Types
Size Diameter
0.5 µl 0.103 mm
1 0.146
2 0.206
5 0.3257
10 0.485
25 0.729
50 1.03
100 1.457
250 2.304
500 3.256
1. ml 4.608 mm
1.25 5.151
2.5 7.285
5 10.3
10 14.567
25 23.033
50 32.573
100 32.573
Cadence Science, Inc.
Formerly Popper & Sons
MICRO-MATE Glass
Size Diameter
0.25 ml 3.47 mm
0.5 3.62
1 4.82
2 8.91
3 8.91
5 11.71
10 14.65
20 19.56
30 22.7
50 28.02
100 35.7
Becton Dickinson
Glass-All Types
Size Diameter
0.5 ml 4.64 mm
1 4.64
2.5 8.66
5 11.86
10 14.34
20 19.13
30 22.7
50 28.6
100 34.9
Sherwood–Monoject
Plastic
Size Diameter
1. ml 4.65 mm
3 8.94
6 12.70
12 15.90
20 20.40
35 23.80
60 26.60
140 38.40
Harvard Apparatus PHD ULTRA
™
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
89
A
ppendices